1-36
1900-009-002 REV B
En
gli
sh
Service
HYDRAULIC FLUID LEVEL CHECK (HYDRAULIC BASE)
TOOLS REQUIRED:
•
(2) Bungee Cords (or equivalent)
•
3/4 in. Socket
•
3/8 in. Drive Ratchet
•
Mobil Aero HFA Hydraulic Fluid (Stryker p/n: 2020-070-475)
PROCEDURE:
1. Raise the rails to the highest position and lower the litter to the lowest height. Apply the brakes.
2. Ensure there are no hydraulic fluid leaks. If leaks are present, replace the jack
(see page 1-35)
.
3. Remove the bottom part of the bellows from the base hood and support them using bungee cords (or equivalent)
to gain access to the fill plug.
4. Using a 3/4 in. socket, slowly turn the fill plug (located on the side of the reservoir) counter-clockwise to allow
excess system pressure to vent. Remove the fill plug.
5. The hydraulic fluid should be visible at the bottom of the hole. If the fluid cannot be seen, add Mobil Aero HFA
hydraulic fluid (p/n 2020-070-475) until the fluid level reaches the bottom of the hole. Replace the fill plug.
CAUTION
Using hydraulic fluids other than the recommended Mobil Aero HFA may cause damage to the jacks.
6. Verify proper jack operation before reinstalling the bellows and returning the stretcher to service.
ITEM
PART NO.
PART NAME
A
45−966
O−Ring
B
45−967
O−Ring
C
390−2−134
Conical Comp. Spring
D
715−1−301
Base Plug
E
715−1−309
Valve Plug
F
715−1−341
Poppet
H
715−100−325
Pump Replacement Assembly
J
926−20−153
Check Valve
K
926−20−154
Seal
L
926−20−156
Seal
M
2025−700−26
Valve Assembly
N
5050−70−50
P.C. Valve
Figure 11