W
ASH FUN
CTION
Wash function
How to access and adjust the wash function.
Step 1: Access settings menu
Step 2: Adjust wash settings
Step 3: Start pump to perform
wash
= Wash function active
= Wash function inactive
= Wash settings sub-menu
= Increase pressure/flow/duration settings
= Decrease pressure/flow/duration settings
= Exit settings menu
5
Wash function
The wash function allows for a customized, quick increase in pressure and flow to
clear debris, control bleeding and improve visualization in the joint. Can customize
to last for a duration up to 120 seconds.
Understanding wash
EXAMPLE:
50% Pressure increase of 50 mmHg
set pressure would become 75mmHg during
wash cycle.
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use.
6
= Pump customization menu
Summary of Contents for CrossFlow
Page 1: ...Doctor 1 Dynamic Medium XF2 STOP CrossFlow quick start guide ...
Page 2: ......
Page 4: ......