100259 REV C
www.stryker.com
Care•Free Companion
®
Pump
CF300
FUNCTIONAL CHECK INSTRUCTIONS
Page 2 of 2
3. Check that all 4 feet are firmly attached to the pump.
4. Check power cord and plug for damage.
5. Tip the pump in several directions and listen for any indication of loose parts. Check housings for damage.
If there is any indication of loose parts or a damaged housing, discontinue use and contact your local dealer or
Stryker’s Customer Service.
PRESSURE CHECK
1. Plug in the pump and turn it on. Allow the pump to operate for at least 60 minutes before proceeding.
2. Screw the pump’s inflator cap onto the tester. Attach
3
/
16
” ID tubing from the manometer to the fitting on the
tester (See Figure 1).
3. Plug the orifice on the end of the tester with a finger.
Repeat a minimum of three (3) times. With the orifice plugged, the pressure reading on the manometer should
be 18-32 mm Hg.
4. With the finger removed from the orifice, the pressure reading on the manometer should be at least 13 mm Hg.
If any of the pressure readings are out of specification, discontinue use and contact your local dealer or Stryker’s
Customer Service.
CURRENT LEAKAGE
The Care•Free Companion Pump has double insulated construction. It is not necessary to measure current
leakage. However, if you choose to do so, refer to UL2601-1 “Medical Electrical Equipment, Part 1: General
Requirements for Safety”, Sub-clause 19.4g.
GROUND RESISTANCE
Double insulated appliances are permitted to have two conductor cords (refer to NFPA 99 “Standard for Health
Care Facilities,” paragraph 9-2.1.2.2). Measuring ground resistance is not applicable to the Care•Free Companion
Pump.
CLEANING
After completing the Functional Check, clean the pump housing with a mild detergent and damp cloth. Wipe dry
with a clean cloth before operating. Do not submerge in water. Do not autoclave.
CONTACT INFORMATION
Contact Stryker Customer Service or Technical Support at: (800) 327-0770 or (269) 324-6500.
Manufactured For
Stryker Medical
3800 E. Centre Ave.
Portage, Michigan 49002
USA
Care•F ree Companion Pump
3/16” ID Flexible T ubing
Manometer
Tester
Orifice
Initiator Cap
Figure 1—Pressure Test Set-up