![Stryker Medical COMPANION Operation And Maintenance Manual Download Page 17](http://html.mh-extra.com/html/stryker-medical/companion/companion_operation-and-maintenance-manual_1386141017.webp)
16
Companion II with T-Band Assembly Instructions
Required Tool
#2 Phillips Head Screwdriver
Required Parts
ITEM
PART NUMBER
PART NAME
QTY.
A
3100-100-045
T-Band Top
1
B
37-21 1
Plastic Cap
1
C
3100-100-085
Split Top Release Handle
1
D
3100-100-078
Gear Rack
2
E
23-69
Pan Hd. Sheet Metal Screw
6
F
3100-100-090
Slide with Clinch Nut
2
G
3100-100-010
Split Top Vanity Assembly
1
H
23-306
Pan Hd. Sheet Metal Screw
8
J
3100-100-079 Gear
2
K
3100-100-080
Blow Molded Tray
1
L
3100-100-095
Split Top Weldment
1
M
7-66
Phillips Truss Hd. Screw
4
N
14-67
Flat Washer
4
Assembly Method
1.0 Place item A (T-band top) upside down on the assembly surface.
1.5 Place item B (plastic cap) on item C (release handle) locking lever as shown on page 14. Position item
D (gear rack) so the gear rack teeth are not visible. Place item C (release handle) into the slot on item
D with the locking lever on item C facing downward through the opening. The paddles on item C should
be opposite the gear teeth on the gear rack.
2.0 While holding item C in the slot in item D, rotate both items so the gear teeth on item D are visible. Place
item D onto item A (T-band top) so the gear teeth on item D are facing toward the center of item A. Align
item D with the holes on item A. Using a #2 Phillips screwdriver, and (2) item E (pan head sheet metal
screw) secure item D to item A. Repeat at the other end of item A (without item C).
NOTE
The gear teeth on item D (gear rack) are facing toward the center of item A. Verify the paddles on item C
are facing toward the outside edges of item A (refer to the drawing on page 14).
2.5 Spray silicon lubricant on the track surface of each item D (gear rack). Wipe off excess lubricant.
3.0 Slide item K (secondary tray) onto the tracks on item D so the molded gear racks are visible. Use the
arrow indicator on item K to center item K over the top of item A.
3.5 Place (2) item F (slide with clinch nut) onto item D. Position item F so the clinch nuts are positioned toward
the top of the assembly and the slot opening is toward item D. Align the slot openings with the through
holes on item D. Insert (4) item H (pan head sheet metal screw) through item F and item D and into item
A. Securely fasten item H with a screwdriver.
4.0 Turn item G (split top vanity assembly) upside down and slide item G onto item K (secondary tray) as
shown in the drawing on page 14. Slide item G into item K until it securely snaps into place.
4.5 Place (2) item J (gear) into item D and item K. Position item J so the gear tooth on item J is at the arrow
indicator of item K. Position item K so item J will fit between both racks. Reference the drawings on
page 14 and 15 if assistance is required.
5.0 Position item L (split top weldment) so the attachment gusset is on the same side as the split top release
handle. Align the slot openings on item L with the clinch nuts on item F (slide). Attach item L to item
F using a Phillips head screwdriver and (4) item M (flat head Phillips screw).