Companion II Column Assembly Instructions
Required Tools:
5/16” Nut Driver
Optional: Rubber Mallet
Required Parts:
ITEM
PART NUMBER
PART NAME
QTY.
A
3100-100-021
Outer Column
1
B
3100-100-032
Outer Tube Friction Grommet
1
C
3100-100-012
Bellows
1
D
3100-100-020
Inner Column Weldment
1
E
3100-100-017
Inner Tube Friction Grommet
1
F
26-262
Clevis Pin
1
G
27-19
Cotter Pin
2
H
3100-100-210
Lift/Lock Assembly
1
J
26-263
Clevis Pin
1
K
3100-100-034
Release Handle
1
L
16-23
Nylock Hex Nut
1
M
52-801
Nyliner Sleeve
2
N
34-267
Nylon Flanged Bushing
1
P
3100-100-65
Outer Shim (.040)
1
3100-100-66
Outer Shim (.050)
1
Q
3100-100-67
Inner Shim (.040)
1
3100-100-68
Inner Shim (.050)
1
Assembly Method
1.0 Slide item C (bellows) over item D (inner column) as shown on the drawing on page 12.
1.5 Insert and snap lock item B (outer tube friction grommet) onto the end of item A (outer column). Use
a rubber mallet, if necessary.
2.0 Slide item D (inner column) into item A (outer column) so item B (friction grommet) rests flush against
the base column plate of item D. Insert item E (inner tube friction grommet) into the open end of item
A (outer column).
2.5 Pull item D (inner column) out of item A (outer column) approximately 6 inches. Insert item H (lift/lock
assembly, single top) into item A with the cylinder shaft end first. Align the T-handle with the holes in
item D (inner column). Insert item F (clevis pin) through both column walls and the T-handle. Fasten
item G (cotter pin) through the slot on item F (clevis pin). Extend item A (outer column) and properly
align the column through holes with the T-handle. Insert item J (clevis pin) through both column walls
and the T-handle. Fasten item G (cotter pin) through the slot in item J (clevis pin).
3.0 Insert the inner column end of the assembly into a holding fixture. Snap (2) item M (nyliner) onto item
K (release handle) as shown on page 12. Align item K (release handle) with item A (outer column) so
the two nyliners nest into the notches on item A.
NOTE
Be sure item K (release handle) is aligned properly with item D (inner column).
3.5 Fasten item L (hex nut) to the release rod and firmly tighten against item K (release handle).
Adjustment Procedure
Condition 1
Condition 2
If table is binding or is hard to lower when the
If table will not lower but doesn’t bind when the
release lever (K) is actuated:
release lever (K) is actuated:
1. remove top assembly
1. remove top assembly
2. adjust nut (L) 1/2 turn counterclockwise
2. adjust nut (L) 1/2 turn clockwise
3. reassemble top to column
3. reassemble top to bottom
4. actuate lever (K).
4. actuate lever (K).
If problem persists, repeat adjustment procedure.
If problem persists, repeat adjustment procedure.