CAUTION
The operator must be fully instructed in how to handle and use electrolytes with
this machine.
CAUTION
The machine is designed to be used with electrolytes recommended by Struers.
Electrolytes that are not recommended by Struers can be dangerous to the
operator or harm the machine.
Procedure
1.
Open the door of the electrolyte compartment.
–
If there is no electrolyte cartridge in the
compartment, push the electrolyte tubes to the
right side of the compartment.
–
If there is an electrolyte cartridge in the
compartment, press the connectors to
disconnect them from the electrolyte cartridge,
and remove the electrolyte cartridge from the
compartment.
Wipe the electrolyte cartridge and connectors
with a damp cloth when you have removed the
used electrolyte cartridge.
2.
Slide the new electrolyte cartridge into the
compartment.
3.
Press the tube over the color coded fitting on the
electrolyte cartridge. When you hear a click, this
indicates that the fitting is secure.
The tube will not click onto the fitting unless the
connector is fully open. To open the connector:
–
Press the metal tab in fully, until you hear a
click.
The connections are color coded:
–
Red
Vent connection and spill return from the pistol
holder.
–
Orange
For electrolyte solution returning from the
polishing chamber.
–
Black
Suction to the polishing chamber.
4.
Close the door of the electrolyte compartment.
A
Red
B
Orange
C
Black