6.1 Safety checks
We recommend carrying out safety checks at least once a year.
Note:
To document the result of the safety check use a test report, see Section 8. Appendix.
XENON NOVA
®
300
MODEL 20
1340 20
6-2
06.11 / V1.6
Work to be carried out
Remark
쐽
Visual inspection
– Housing and accessories
– Inscriptions, manufacturer’s identification
plate data
– CE mark, KARL STORZ inspection label
– Instruction manual
– Power fuses
5)
– Fuse label
쐽
쐽
Test for proper operation
– Power switch
– Power supply
– Lamp replacement
– Light cable
– Manual intensity control
– Fans
쐽
쐽
Electric safety measurements
(IEC 62353)
– Protective ground resistance:
(with power cord)
울
0.3
액
– Earth leakage current:
울
1.00 mA
– Touch current:
울
0.50 mA
– Patient leakage current:
울
0.05 mA
– Patient leakage current:
울
0.05 mA
(line voltage on the applied part)
no external damage
correct, legible, clean, wipeable, securely attached
fixed to the housing
present
correct ratings, undamaged, securely positioned
applied next to fuse holder, correct fuse ratings
functions perfectly within the stated voltage range
(100 VAC … 125 VAC / 220 VAC … 240 VAC)
easy to perform
can be inserted easily into the corresponding
socket, engages easily into detented position
brightness adjustment knob
5) For correct fuse ratings see Section 3.4 Technical data of the XENON NOVA
®
300.
Summary of Contents for 201340 20
Page 8: ......
Page 10: ......
Page 16: ......
Page 18: ......
Page 25: ...XENON NOVA 300 MODEL 201340 20 06 11 V1 6 3 7 3 2 3 1 Circuit diagram of the lamp timer board...
Page 30: ......
Page 38: ...4 9 3 Figure 2 XENON NOVA 300 MODEL 201340 20 4 8 4 8 06 11 V1 6...
Page 39: ...4 9 4 Figure 3 XENON NOVA 300 MODEL 201340 20 06 11 V1 6 4 9...
Page 41: ......
Page 45: ......
Page 47: ......
Page 53: ......
Page 55: ......
Page 57: ......
Page 58: ......