background image

Daily Maintenance

Clean the door seal using a lint free cloth and soapy water.
Check the door seal for any signs of damage and replace if necessary.
Clean the face of the chamber that the door seal closes onto.
Drain both reservoirs at the end of each day and leave empty overnight.

Daily Testing

Steam penetration test - put a TST test strip in a Helix tester. Put the Helix tester in the steriliser
and run a vacuum cycle. The test is passed if the TST test strip changes colour.
Automatic control test - run a cycle and observe. The test will be considered satisfactory if the

following requirements are met:

 the steriliser does not error at any point during the cycle

 the door cannot be opened until the cycle is complete

 a visual display of “cycle complete” occurs at the end of the cycle

 no mechanical or other anomaly is observed

Weekly Testing

Run the automatic air leakage test. This test will show that the steriliser is capable of pulling and
holding a certain level of vacuum. The display will indicate a pass or fail result at the end of the
test cycle. The test cycle is accessed from stand-by with the door open - press the “

i

” button, then

the “Options” button.

Servicing

Arrange to have the steriliser serviced by a trained service engineer every six months.

Annual Inspection

In order to comply with Health and Safety legislation you should arrange to have your steriliser
inspected for pressure safety  by a “competent person” once a year.

Annual Testing (or Validation)

In  order  to  check  that  all  cycle  parameters  are  being  met  you  should  arrange to have your
steriliser tested annually by a trained service engineer.

4

Do not permit any
person other than
qualified personnel
to supply parts for,
service or repair your
Acuatic Vacuum
steriliser.

Acuatic Vacuum steriliser

Care and Data Logger

Only use chlorine free
cleaning products.

Summary of Contents for Acuatic Vacuum 12 B

Page 1: ...Sterling Sterilisers Acuatic Vacuum 12 B User Instructions...

Page 2: ...n EC No 1882 2003 0086 indicates that the Notified Body employed to check conformation with the above is BSI This symbol indicates that the number next to it is the serial number of the steriliser Thi...

Page 3: ...Logger Daily Maintenance 4 DailyTesting 4 Weekly Testing 4 Servicing 4 Annual Inspection 4 Annual Testing 4 Getting Started GettingStarted 5 Filling 5 Water Quality 5 Draining 5 Options 5 Operation S...

Page 4: ...onstruction structural parts are manufactured from stainless steel or aluminium Impossible to overheat innovative engineering eliminates downtime due to overheating Easy water fill just pour into the...

Page 5: ...ising Drying 138 C 134 C 1 0 2 Pressure bar 1 Time Air removal and heating Drying Sterilising B type A door seal B chamber with tray rack and trays C water filling spout D touch screen display E clean...

Page 6: ...d instruments the internal surfaces of the steriliser and the steriliser door may be hot Do not touch these parts until cooling has taken place and exercise care when handling just sterilised instrume...

Page 7: ...is observed Weekly Testing Run the automatic air leakage test This test will show that the steriliser is capable of pulling and holding a certain level of vacuum The display will indicate a pass or f...

Page 8: ...this is not recommended because of the risk of the heaters becoming scaled Draining When the used water reservoir is full the display will ask for it to be drained This will need to be done before the...

Page 9: ...If the N type cycle is enabled you will be offered a choice of starting N type or B type cycles Processing When a cycle is running the display will show the cycle time elapsed the operating temperatu...

Page 10: ...ep the Add shortcut to desktop checkbox ticked To import cycle data from the SD memory card to your computer run the Securilog application by double clicking the Securilog icon on your desktop Remove...

Page 11: ...with the Medical Devices Directive 4 A Product Registration and Customer Survey form Please take the time to print this out complete it and return it to Sterling Sterilisers so that we can keep our re...

Page 12: ...apped pouched or not Limits Design pressure 1 to 3 bar Design temperature 2 to 144 C Water Water used per cycle 200 to 700 ml Clean water reservoir volume 2 litres Used water reservoir volume 2 litres...

Reviews:

Related manuals for Acuatic Vacuum 12 B