MAINTENANCE

REPLACING YOUR AC CARTRIDGES 

After the Ac has notified you that a cartridge replacement is necessary either by the Water Quality Alarm LED 

light on the faceplate turning solid red and/or after hearing the Ac’s audible alarm (beeping), 

DO NOT

 replace 

all three AC-CXL or AC-C cartridge at once (unless directed to do so by a Sterisil

®

 certified technician). Follow 

these steps to replace your Ac cartridges:

WATER QUALITY

ALARM

Autoclave Water

       Start date

European A.R.
Emergo Europe
Molenstraat 15
2513 BH The Hague
The Netherlands

 Sterisil, Inc.

835 S. Hwy 105 Ste. D

Palmer Lake, CO USA 80133

+1 719 622 7200

0120

Contact Sterisil to receive a

free recycling label.

GO GREEN WITH 

STERISIL!

Ac

DEIONIZATION

CARTRIDGE

P/N AC-C

Autoclave Water

        Start date

European A.R.
Emergo Europe
Molenstraat 15
2513 BH The Hague
The Netherlands

 Sterisil, Inc.

835 S. Hwy 105 Ste. D

Palmer Lake, CO USA 80133

+1 719 622 7200

0120

Contact Sterisil to receive a

free recycling label.

GO GREEN WITH 

STERISIL!

Ac

DEIONIZATION

CARTRIDGE

P/N AC-C

Autoclave Water

       Start date

European A.R.
Emergo Europe
Molenstraat 15
2513 BH The Hague
The Netherlands

 Sterisil, Inc.

835 S. Hwy 105 Ste. D

Palmer Lake, CO USA 80133

+1 719 622 7200

0120

Contact Sterisil to receive a

free recycling label.

GO GREEN WITH 

STERISIL!

Ac

DEIONIZATION

CARTRIDGE

P/N AC-C

Autoclave Water

DIRECTIONS FOR USE

BEEPING OR SOLID RED LIGHT

-Follow these steps to rotate in new Ac cartridges

o  Move third cartridge to farthest left position.

  o Install new cartridges in second and third positions.

CHECKING THE Ac SYSTEM STATUS (WATER HARDNESS)

 -Flip white switch to Source or left position. Record reading

 -Flip white switch to DI or right position. Record reading.  This number should be less than 10.

o if DI is greater than ten, refer to the directions above for BEEPING OR SOLID RED LIGHT.

1.  From left to right, remove 

the first two cartridges and 

dispose of them normally. 

2.  Move the remaining 

cartridge from the third 

cartridge position to the 

first cartridge position. 

3.  Install two new Ac 

cartridges in the second 

and third cartridge 

positions.

WATER QUALITY

ALARM

Autoclave Water

        Start date

European A.R.
Emergo Europe
Molenstraat 15
2513 BH The Hague
The Netherlands

 Sterisil, Inc.

835 S. Hwy 105 Ste. D

Palmer Lake, CO USA 80133

+1 719 622 7200

0120

Contact Sterisil to receive a

free recycling label.

GO GREEN WITH 

STERISIL!

Ac

DEIONIZATION

CARTRIDGE

P/N AC-C

Autoclave Water

DIRECTIONS FOR USE

BEEPING OR SOLID RED LIGHT

-Follow these steps to rotate in new Ac cartridges

o  F

o  Move third cartridge to farthest left position.

  o Install new cartridges in second and third positions.

CHECKING THE Ac SYSTEM STATUS (WATER HARDNESS)

 -Flip white switch to Source or left position. Record reading

 -Flip white switch to DI or right position. Record reading.  This number should be less than 10.

o if DI is greater than ten, refer to the directions above for BEEPING OR SOLID RED LIGHT.

Autoclave Water

        Start date

European A.R.
Emergo Europe
Molenstraat 15
2513 BH The Hague
The Netherlands

 Sterisil, Inc.

835 S. Hwy 105 Ste. D

Palmer Lake, CO USA 80133

+1 719 622 7200

0120

Contact Sterisil to receive a

free recycling label.

GO GREEN WITH 

STERISIL!

Ac

DEIONIZATION

CARTRIDGE

P/N AC-C

Autoclave Water

        Start date

European A.R.
Emergo Europe
Molenstraat 15
2513 BH The Hague
The Netherlands

 Sterisil, Inc.

835 S. Hwy 105 Ste. D

Palmer Lake, CO USA 80133

+1 719 622 7200

0120

Contact Sterisil to receive a

free recycling label.

GO GREEN WITH 

STERISIL!

Ac

DEIONIZATION

CARTRIDGE

P/N AC-C

WATER QUALITY

ALARM

Autoclave Water

       Start date

European A.R.
Emergo Europe
Molenstraat 15
2513 BH The Hague
The Netherlands

 Sterisil, Inc.

835 S. Hwy 105 Ste. D

Palmer Lake, CO USA 80133

+1 719 622 7200

0120

Contact Sterisil to receive a

free recycling label.

GO GREEN WITH 

STERISIL!

Ac

DEIONIZATION

CARTRIDGE

P/N AC-C

Autoclave Water

DIRECTIONS FOR USE

BEEPING OR SOLID RED LIGHT

-Follow these steps to rotate in new Ac cartridges

o  F

o  Move third cartridge to farthest left position.

  o Install new cartridges in second and third positions.

CHECKING THE Ac SYSTEM STATUS (WATER HARDNESS)

 -Flip white switch to Source or left position. Record reading

 -Flip white switch to DI or right position. Record reading.  This number should be less than 10.

o if DI is greater than ten, refer to the directions above for BEEPING OR SOLID RED LIGHT.

Summary of Contents for AC

Page 1: ...719 622 7200 Sterisil com STERISIL AC SYSTEM INSTALLATION AND OPERATING MANUAL ...

Page 2: ...D HELP OR HAVE QUESTIONS Our technical support staff is available and ready to help Please call email or visit us online for additional support sterisil com TABLE OF CONTENTS Introduction 3 Sterisil Ac Features 4 Installation Details 5 Installation Instructions 6 Maintenance 11 Troubleshooting and FAQ 13 ...

Page 3: ...with 10 ppm Total Dissolved Solids TDS Audible and visual alarms make cartridge replacement worry free and the Sterisil Ac requires only minutes of maintenance Its space saving compact design allows for easy installation even under sinks This manual will show you how to install and maintain your new Sterisil Ac unit and answer a few frequently asked questions that you may have For additional suppo...

Page 4: ...ECTIVE Produces distilled quality water with 10 ppm TDS for use with table top autoclaves Water Quality Monitor samples the water purity in real time Audible and visual alarms alert you when cartridges need to be replaced Dispensing faucet with LED monitoring light Backup filters supply your office with water for your table top autoclaves even after audible and visual alarms trigger Designed for o...

Page 5: ...t typically include a sharp razor knife or tube cutter measuring tape drill assorted bits and screws screwdrivers wrenches level and clean rags LOCATION Generally the Ac System is installed centrally within the dental office in the equipment room sterilization center hallway or laboratory If the hinged mounting bracket is selected the typical mounting location would be on the wall near the source ...

Page 6: ... plate at the desired location by placing it on the wall b Ensure that the mounting plate is level Then mark each of the holes indicated below with a pen or pencil c Using a drill bit drill holes at the points you marked in step b d Insert the drywall anchors into the holes and gently hammer them all the way in e Place the mounting plate over the holes and using the Phillips 10 x 1 sheet metal scr...

Page 7: ...cet assembly into the hole as shown c Tighten faucet in place and install quick connect fitting 6 Establish Sterisil Ac plumbing connections a Ensure that the municipal cold water supply line to your sink is turned off From the municipal cold water supply line install a tee valve for under sink installation supplied by customer and reduce to via either quick connect fitting or compression fitting ...

Page 8: ...ater supply to the Ac unit and cut the tubing leaving enough slack to allow the system to slide out from the cabinet Firmly insert the tubing into the red port labeled SOURCE WATER IN on the side of the unit d Run another length of white tubing to the quick connect fitting on the dispensing faucet Cut the tubing leaving enough slack to allow the system to slide out from the cabinet ...

Page 9: ... turn on and the alarm may sound after initially plugging in the unit If so unplug the power cord and wait to plug in until after cartridges are installed and water is turned on to the unit b b Connect the RJ12 data line from the Ac unit to the data union included with the unit Connect the other end of the data union to the RJ12 data line from the faucet ...

Page 10: ...10 GP6002 B AC SYSTEM DIAGRAM ...

Page 11: ...water TDS reading To do this turn on the blue water quality monitor on the faceplate of the Sterisil Ac by pressing the silver on off button in the upper right hand corner of the monitor Ensure that the white slider is to the left next to Source Record the number now displayed in the LCD window NOTE this number may change while you are attempting to take the reading simply record the highest numbe...

Page 12: ...and third cartridge positions WATER QUALITY ALARM Autoclave Water Start date European A R Emergo Europe Molenstraat 15 2513 BH The Hague The Netherlands Sterisil Inc 835 S Hwy 105 Ste D Palmer Lake CO USA 80133 1 719 622 7200 0120 Contact Sterisil to receive a free recycling label GO GREEN WITH STERISIL Ac DEIONIZATION CARTRIDGE P N AC C Autoclave Water DIRECTIONS FOR USE BEEPING OR SOLID RED LIGH...

Page 13: ... Ac unit is turned on If the cold water supply is not turned on turn it on and check for water again If the cold water supply is on continue troubleshooting 1 Taking a new unused Ac CXL or Ac C cartridge remove the first cartridge and screw the unused cartridge in to the first position Check for water flow If water flow returns dispose of the used cartridge and follow the steps under Replacing You...

Page 14: ...Inc 835 S Hwy 105 Ste D Palmer Lake CO USA 80133 THANK YOU Thank you for purchasing your new Sterisil Ac System We know that you your patients and staff will be pleased with this Unit Should you have any installation questions please contact us at 719 622 7200 or online at Sterisil com ...

Reviews: