4-34
920011-649
Operator Manual
Processing Instructions
4.8.4 Low Chemical Warning
This option is used to warn operator when level of chemical in Bag-
in-a-Box indicated in display message is getting low. It should
prevent having a chemical empty alarm during a cycle.
Low chemical warning is enabled in
Miscellaneous Values
menu of
Supervisor mode by setting
CHEMICAL WARNING option
to
YES
(refer to
).
Level of chemical in Klenzyme
®
Enzymatic Presoak and Cleaner
Bag-in-a-Box is automatically verified during start-up of SINGLE and
DOUBLE cycles with WASHING PHASE. Level of chemical in CIP
200
®
Acid-Based Process and Research Cleaner Bag-in-a-Box is
automatically verified during start-up of D-LONG cycle.
1. If chemical level is low when
START
touch pad is pressed,
display shows:
or:
alternating with:
and:
IMPORTANT: When previous messages are displayed,
approximately five cycles can be processed before indicated
Bag-in-a-Box is completely empty. "FAULT 08 or 09, CYCLE
DELAYED" alarm is triggered, indicating a Bag-in-a-Box is empty.
2. Replace indicated Bag-in-a-Box as explained in
.
or:
Press
STOP
touch pad to ignore message and process selected
cycle.
KLENZYME
IS LOW
CIP 200
IS LOW
REPLACE AND
PRESS START
PRESS STOP
TO IGNORE
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...