background image

n

Figure 10

SMO-type impeller with openings that ex-

pose the bearing surfaces and weld plate during the clean-
ing cycle.

CORROSIVE AGENTS HAZARD:

Many cleaning

agents are corrosive. Please exercise adequate care

when handling.

4.5.2. Cleaning guidance for SMO/SMMS/Sanimixer-

type impeller

This section provides general guidelines for CIP proce-

dures for Sterimixer with SMO style impellers, Sterimixer
SMMS and Sanimixer.

The bearing assembly on a SMO/SMMS/Sanimixer-

type impeller head is exposed. When directing flow from
a spray ball to the impeller head, the bearing surfaces will
get in contact with the CIP fluid.

The openings on top of the SMO-type impeller head

also allow the CIP fluid to hit the weld-plate under the im-

peller, please refer to Figure 10. In order to hit all surfaces,

the impeller head needs to rotate at low speed during the
CIP cycle (e.g. 100 rpm).

The cleaning procedure can be divided in two steps:

4.5.3. Step 1: CIP solution cleaning

Start CIP solution flow through the sprayball at the rec-
ommended flow rate. Open the tank outlet valve com-

pletely to allow full drainage. Operate mixer at low speed
(around 50-100 rpm).

Ensure that the sprayball flow is hitting the impeller

head. Typically, coverage test with riboflavin (usually

performed during FAT) will provide evidence that the

impeller head is fully wetted by the cleaning solution.

The duration of the cycle depends on the application

and requires to be validated by swabbing vessel and mixer
surfaces.

4.5.4. Step 2: WFI (water for injection) flushing

Repeat step 1 using WFI (water for injection) through the
sprayball.

4.6. Guidance for sterilizing Sterimixer/Sanimixer

The Sterimixer/Sanimixer are designed for optimum per-

formance in steaming in place (SIP) without any need for
dismounting. The selection of method from those listed
below depends on the individual conditions.

CAUTION:

The permanent magnets in the impeller

and the rotor withstand temperatures up to +225

C.

Never expose these parts to higher temperatures. The
drive unit and gearbox shall not be exposed to ambient
temperatures above +40

C.

+

NOTICE:

Use caution when exposing to temperatures

over +135

C

as this may decrease the life expectancy

of the elastomeres used in the bearing assembly.

A certain delay in the rise of temperature between the

inside of the impeller and the vessel is caused by the mass
of the impeller head. This delay is influenced by:

• Sterimixer/Sanimixer size: the larger the Sterim-

ixer/Sanimixer, the longer the delay.

• Sterilization method: steam sterilization gives the

shortest delay. Hot air sterilization gives the longest
delay.

When evaluating sterilization temperature one should

focus on the location that is most likely to have the lowest
temperature (typically this would be on the inside, in the
impeller cavity). A thermal element should be used if the
system is sterilized in a hot air oven or autoclave. For
applications with SIP/SHW where the impeller rotates,
the validation should be carried out by biological tests.

4.6.1. Steaming in place (SIP)

Example of SIP procedure:

• To remove any condensation,the impeller may be ro-

tated at a low speed (i.e. 50 rpm) for a few seconds,
approximately every five minutes.

• When the sterilization temperature in the vessel has

been achieved, the periodic rotation of the Sterim-
ixer/Sanimixer shall be stopped.

4.6.2. Super heated water sterilization in place (SHW)

The vessel and the mounted drive unit are sterilized with

super heated water. Example of SHW procedure:

• The vessel is filled with WFI (water for injection) to a

suitable level.

• During heating, sterilization and cooling the Sterim-

ixer/Sanimixer should rotate.

• Vortices (suction whirlpools) must not arise. The wa-

ter level and speed control this.

• During the emptying phase the Sterimixer/Sanimixer

must not rotate.

4.6.3. Sterilization in autoclave or hot air oven

Dismount the drive unit as per section 3.4.2.

CAUTION:

The vessel and the impeller (without the

drive unit) must be sterilized in an upright position
in an autoclave or hot air oven to prevent the impeller
sliding off the bearing. To prevent this, a magnetic drive
dummy can be installed to avoid any risk of damage
and hold the impeller in place.

10

|

© Steridose

www.steridose.com

Summary of Contents for Sanimixer

Page 1: ...4 Mounting and dismounting the drive unit 5 3 5 Connection to main power 6 3 6 Installation check and calibration 6 4 Operation 7 4 1 Allowed operating temperature range 7 4 2 Allowed pressure range 7 4 3 Chemical compatibility 7 4 4 Operational limits 7 4 5 Further operational advice 9 4 6 Guidance for sterilizing Sterim ixer Sanimixer 10 4 7 Specific information for the use of Sterim ixer Sanimi...

Page 2: ... or indirectly create risks to the safety and health of the person undertaking the work or the Sterim ixer Sanimixer and or its components while the work is being undertaken It is the responsibility of the user to ensure that appro priate controls and precautions are identified and applied in relation to the work covered by this document in ac cordance with relevant statutory legal and industry re...

Page 3: ...er is not to be in stalled immediately the Sterimixer Sanimixer should be repacked in the original package and stored in a suit n Figure 1 Mounting the bearing unit on the weld plate in the vessel The figure shows the bearing unit and the PTFE washer Use a torque wrench refer to table 2 for recommended torque able storage after receipt and inspection The following should be considered with regards...

Page 4: ...ve unit and the magnet rotor are installed Failure to do so will cause a sudden attraction between the mounted rotor and the impeller causing the latter to violently hit the bearing assembly with likely bearing damage to occur CAUTION Before mounting the impeller make sure that the drive unit is NOT mounted on the vessel NOTICE To facilitate mounting and dismounting the impeller Steridose has deve...

Page 5: ... driven gearbox flange and rotor some differences occur depending on type of drive and size Be certain that you can handle the weight of the drive unit Take a firm grip of the gearbox to balance the weight of the drive unit and to minimize the risk of damage caused by crushing when dismounting Use appropri ate lifting device for the larger models SM 120 190 and larger 3 4 1 Mounting the drive unit...

Page 6: ...For AC drive applications The Sterim ixer Sanimixer starts at low speed 5 Visually check the direction of rotation for the im peller which should be clockwise when seen from above If the Sterimixer Sanimixer rotates counter clockwise refer to section 6 6 Stop the Sterimixer Sanimixer immediately after completed verification CAUTION Dry operation of the Sterimixer Sanimixer will damage the bearings...

Page 7: ...Allowed pressure range Weld plate size Design pressure range All sizes 1 7 bar g 14 5 101 5 psig 4 3 Chemical compatibility The impeller and bearing are not to be exposed to strong chemicals Nominal pH range for tungsten carbide TC n Figure 6 Sterimixer Sanimixer should never be allowed to run dry n Figure 7 The bearing lubrication principle of the SMA type impeller bearings is 2 14 but silicon ca...

Page 8: ...s IEC ATEX NEMA XP NEMA Stainless Steel etc this may vary Bearing materiala Tungsten Carbide Silicion Carbide Dri amond Chemical compatibility pH range 2 14b Best inert Same as silicon carbide Ease of installation and handling Best Harder than tungsten car bide but more brittle Somewhat better than sili con carbide Survivability under dry running conditions Good Poor Best lowest coefficient of fri...

Page 9: ...optimum performance in cleaning in place CIP without any need for dismounting n Figure 9 Succesful cleaning of the impeller relies on a combination of the self cleaning flow through design of the SMA type impeller the cleaning agent used and respecting the prescribed time and temperature The flow channels in the impeller provide a process fluid flow through the bearings In order to achieve this th...

Page 10: ...e drive unit and gearbox shall not be exposed to ambient temperatures above 40 C NOTICE Use caution when exposing to temperatures over 135 C as this may decrease the life expectancy of the elastomeres used in the bearing assembly A certain delay in the rise of temperature between the inside of the impeller and the vessel is caused by the mass of the impeller head This delay is influenced by Sterim...

Page 11: ...nverter s set value speed exceeds 10 the drive is to be disabled interlocking In this case the magnetic field between the inner and outer magnetic rotor is broken indicating a malfunction 4 7 4 Assessment Components outside of the tank are in Ex zone 1 or 2 and are delivered with the corresponding ATEX declaration of conformity from the manufacturer Ex II 2G b c de IIC T4 The machine mixer tank wi...

Page 12: ... lubri cation creating excessive wear and heat and subse quently premature failure It is therefore important that the operator is properly trained and familiar with all procedures If the Sterimixer Sanimixer is operated beyond its operating limits severe damage might occur subse quently creating a failure It is essential that the oper ator has received specific training together with the full unde...

Page 13: ...e a year Note that the interval will be highly dependant on the application Remove and check the bearing and the impeller for signs of surface damage and general wear of the bear ings Change the bearing and the impeller if the surface is coarse or uneven scratched or if fragments have been knocked out from the bearing The impeller can be sent to Steridose for refurbishment Change the PTFE washer w...

Page 14: ... female bear ing b Worn out bearings c Missing male bearing d Mechanical interference con tact between impeller and weld plate e Mechanical interference con tact between magnet rotor and weld plate a Stop the mixer Do not restart un til the fluid level covers the impeller b Replace the bearing and the im peller c see b d Reinstall the weld plate correctly e Realign the magnet rotor Noise from the ...

Page 15: ...www steridose com Installation Operation 15 ...

Page 16: ...sk for process interference Steridose partners with the best distributors and repre sentatives in the industry all over the world Together we become the perfect mix a premium product with global references combined with local presence for product and application support Steridose AB Himmelsbodavägen 7 SE 147 39 Tumba Sweden Phone 46 8 449 9900 info steridose com www steridose com Steridose Inc 502...

Reviews: