4 Preparing the Equipment for Operation
Operating Instructions
MedAir
, February 2007
4.
Preparing the Equipment for
Operation
4.1.
Setting up the MedAir
The
MedAir
is designed for use in dry areas
free from spray and dripping water. Moist or
wet ambient conditions can greatly impair the
functioning of the equipment.
The ambient temperature may not 40
°C.
Should the relative humidity exceed 70 %, it is
recommended that the equipment only be
operated in air-conditioned areas.
The air intake filter positioned at the rear of the
unit may not be covered by any objects.
The rear side of the unit must be at least 10 cm
from the wall.
4.2.
Connection to the mains
power supply
Check to determine whether the mains power
voltage matches the power voltage as
indicated on the rating plate on the unit.
Make sure that the mains power switch
"ON/OFF is in the OFF position.
Plug the mains power connection cable found
at the rear of the unit into the wall socket-outlet
in the room.