7.
Care and Servicing
7.1. Cleaning
For reasons of hygiene, the unit should
regularly be cleaned with a wipe disinfectant
using a commercially available household
detergent or disinfectant (e.g. Sagrotan).
The intake filter matting and the bacterial
micro-filter must be replaced every 6 months or
after every 500 operating hours.
The intake sound damper should be replaced
when found dirty or at least every 6 months.
7.2.
Safety technical inspections
Safety technical inspections as well as
checking the quality of air generated are to be
carried out annually or after 3500 operating
hours by the manufacturer or by the authorized
Technical Support team of F. STEPHAN.
7.3. Servicing
For reasons of safety, it is recommended that
the servicing of the
MedAirs
be carried out
annually in conjunction with the safety
technical inspections.
Servicing is to be carried out only by the
authorized Technical Support team of F.
STEPHAN.
Only original replacement parts supplied by F.