Stephan 161061000 Operating And Service Instructions Download Page 5

1     General Information 

 

 

Operating Instructions 

MedAir

, February  2007

 

1.7. Technical 

Data 

 
Ambient conditions

 

 

Operation: 

Temperature:   

15 –40 °C 

Operating pressure: 

max. 3.5 bar 

 

Storage: 

Temperature:   

15 – 40 °C 

rel. humidity: 

 

10 – 65 % 

protect against moisture and humidity 

store dust-free and free from frost 

 
Electrical supply

 

 

Mains power:   

 

230V / 50Hz 

Power consumption: 

 

250 W 

 Current 

consumtion: 

 approx. 

5A 

 Power 

output: 

  600 

VA 

 Fuse 

protection:  3.15 

AT 

 

Noise level: 

  

 

45 dB (A) 

 Operating 

pressure: 

 3.5 

bar 

 

Output flow: 

 

0 – 45 l/min 

 

General

 

 

Protection rating: 

 

Inspection / maintenance: 

annually or every 3500 operating hours 

 

Dimensions (L x W x H): 

500x510x1150 

 

 Weight: 

   50 

kg 

 Operating 

position: 

 up-right 

 
 

Summary of Contents for 161061000

Page 1: ...EPHAN GmbH Medizintechnik D 56412 Gackenbach Kirchstr 19 Tel 06439 9125 0 Fax 06439 9125 111 info stephan gmbh com www stephan gmbh com MedAir Mobile Compressed Air Unit h Operating and service Instructions ...

Page 2: ...mbly of the system Should there arise any problems with the functioning of the system please get in touch with the Technical Support team at F Stephan GmbH or the authorized dealer from whom you have received the equipment and received training in its use The manufacturer guarantees the safety and reliability of the MedAirs only when the equipment is operated in compliance with the Operating Instr...

Page 3: ...al terminology 4 1 7 Technical Data 5 2 Safety Instructions 6 2 1 Danger warnings 6 2 2 Caution warnings 6 3 Description of the Design and Functioning 7 3 1 Front view 7 3 2 Side view left 8 3 3 Rear view 8 4 Preparing the Equipment for Operation 9 4 1 Setting up the MedAir 9 4 2 Connection to the mains power supply 9 5 Operation 10 6 Problems Causes and Remedies 11 7 Care and Servicing 12 7 1 Cle...

Page 4: ... cardboard cartoning can be reused or disposed of as waste paper The immediate wrapping around the system is of CFC free material It can be disposed of along with the foil using the recyclable waste bin 1 5 Introduction The German Equipment Regulations MPG the Medical Products Operations Act MPBetreibV and the law governing technical materials stipulate that the attention of the user operator must...

Page 5: ...moisture and humidity store dust free and free from frost Electrical supply Mains power 230V 50Hz Power consumption 250 W Current consumtion approx 5A Power output 600 VA Fuse protection 3 15 AT Noise level 45 dB A Operating pressure 3 5 bar Output flow 0 45 l min General Protection rating I Inspection maintenance annually or every 3500 operating hours Dimensions L x W x H 500x510x1150 Weight 50 k...

Page 6: ...orized personnel The MedAir must be operated in accordance with these Operating Instructions The MedAir may not be operated in areas exposed to a danger of explosion The equipment may not be operated without a micro filter The air intake filter on the rear side may not be covered by any objects In order to ensure an optimal supply of fresh air and air circulation the equipment is only to be operat...

Page 7: ...ter bacterial filter and an additional condensation water extractor The MedAir is mobile vibration and sound insulated A side rail 25x10 mm allows the mounting of additional equipment Monitoring and operation are via a control panel with mains power switch operating hours counter fuses and pressure gauge 3 1 Front view 1 Power switch ON OFF 2 Fuses 3 15 AT 3 Operating hours counter 4 Pressure gaug...

Page 8: ...1 Air intake filter matting 2 Bacterial micro filter with the intake sound damper positioned behind 3 Moisture micro filter with moisture micro filter insert 4 Mains power connection cable 5 Drainage valve Depending on fill level the condensation water is to be drained from catch tank For doing this the drainage valve lever 5 must be pushed upwards 1 1 2 3 4 5 ...

Page 9: ...ty exceed 70 it is recommended that the equipment only be operated in air conditioned areas The air intake filter positioned at the rear of the unit may not be covered by any objects The rear side of the unit must be at least 10 cm from the wall 4 2 Connection to the mains power supply Check to determine whether the mains power voltage matches the power voltage as indicated on the rating plate on ...

Page 10: ... MedAirs is activated which ensures adequate cooling of the unit s interior Generation of compressed air begins approximately 10 seconds after that The compressor of the MedAirs builds up an operating pressure of 3 8 bar in approximately 4 seconds The pressure gauge on the front side of the unit shows the pressure at the AIR outlet The outlet connection coupling is positioned on the left side of t...

Page 11: ... filter is dirty or clogged Leakage in line Defective compressor Gas usage is too high Replace micro filter Contact Technical Support Reduce gas usage Pressure continuously below 3 5 bar Intake sound damper is dirty Defective pressure safety valve Replace intake sound damper Replace defective presurre safety valve Not functioning Mains power failure The equipment automatically restarts after a res...

Page 12: ... replaced when found dirty or at least every 6 months 7 2 Safety technical inspections Safety technical inspections as well as checking the quality of air generated are to be carried out annually or after 3500 operating hours by the manufacturer or by the authorized Technical Support team of F STEPHAN 7 3 Servicing For reasons of safety it is recommended that the servicing of the MedAirs be carrie...

Page 13: ...ng 4 47 x 1 78 for nippel coupling AIR when leakage detected Items 1 to 6 can be ordered in a set Art No 1 610 61 500 which is sufficient for three years The set additionally includes a replacement parts case and diverse small parts Optional accessories Item Article number Article description Quantity 1 1 602 61 112 Connection hose for compressed air length 3 m 1 2 1 600 40 010 Shock proof socket ...

Page 14: ...s a warranty of 24 months beginning at purchase date Modifications or repair work may only be carried out by the manufacturer F Stephan GmbH or authorized technicians Otherwise the warranty becomes void Warranty shall also become void as a result of improper use and treatment of the equipment ...

Page 15: ...Notes Operating Instructions MedAir February 2007 Personal Notes ...