background image

STEINDORFF

®

 

 

5.

 

Orthogonal polarized observation: 

Because the polarizer is located in the light path, this time is single polarized light state after 

focusing. Slide the analyzer into the light path, to adjust the analyzer adjusting rob on the 

“0”position, also the polarizer is located on the“0”position, this time is orthogonal polarized light 

state (polarized direction of the polarizer is at west-east, polarized direction of the analyzer is at 

north-south). 

When using 10×objective or lower. Please remove compensation lens, and adjust condenser to 

proper place; when using 20 × objectives or higher, please push compensation lens to the previous 

position and lift the condenser to the highest position. 

Insert gypsum (λ), mica (1/4λ) sample, quartz wedge into complementary inlet according to your 

need in order to test pixel. 

6.

 

Observation of cone light: 

Observation of cone light is use 20× or higher objectives, push the Bertrand lens and 

compensation lens into the light path, adjusting center of the Bertrand lens, to observe the cone 

light characteristic of the sample. 

 

VII.

 

MAINTENANCE: 

1.

 

If there is any dirt settled on the lens, wipe it off gently with some lens paper. If alcohol is used, be 

careful not to let it penetrate through the lenses and dissolve the gumming. Any dust settled on the 

lenses can be blown away with an air blower or wiped off with a clean soft camel hair brush. 

2.

 

In cleaning mechanical parts and applying non-corrosive lubricant, take special care not to touch 

the optical elements, especially the objective lenses. 

3.

 

Avoid high temperature. 

4.

 

Put objectives and eyepieces into box after using, and keep them in a cool, shady and dry place, 

free from dust, fumes and moisture.  If not stored in its box, cover with a hood to protect from 

dust. 

5.

 

Like other optical instruments, the microscope should be kept out of direct sunlight in a cool, dry 

place, free from dust, fumes and moisture.  It should be stored in a case or covered with a hood to 

protect it from dust. 

6.

 

The microscope has been carefully tested and inspected.  

 

Summary of Contents for NYMC38000

Page 1: ...STEINDORFF NYMC38000 Polarizing Microscope In order to exert performance of this microscope and to ensure the safety please read the operating instruction carefully before use...

Page 2: ...ght and cone light observation and micrography It is collocated gypsum mica 1 4 sample quartz wedge attachable mechanical stage and other attachment It is a set of new product that has maturity functi...

Page 3: ...bjective center Condenser Screw for adjusting condenser center Adjustable graduation ring for polarizer Collector Rotatable mechanical stage Removable holder of compensation lens Fine focusing knob Ob...

Page 4: ...fication Eyepiece 4 10 25 40 63 10 40 100 250 400 630 5 Adjustable center quadruple nosepiece 6 Rotatable mechanical stage diameter 160mm 360 part scale and cursor precision 6 7 Focusing Coaxial coars...

Page 5: ...00 polarizing microscope 1pc 2 Strain free achromatic objective 4 0 10 1each 10 0 25 1each 25 0 40 1each 40 S 0 65 1each 63 S 0 85 1each 3 Eyepiece 10 Wide field eyepiece 2 each 10 Cross eyepiece 1eac...

Page 6: ...ar or trinocular head onto the transmission adapter and tighten with the fixing screw with finger 4 Eyepiece Insert eyepiece into the viewing tube 5 Objective Make the mechanical stage down in order t...

Page 7: ...ides should be prepared with cover slips over the specimen for observation under the 100 objective The 100 oil immersion objectives should be wiped off with a piece of soft clean cloth or lens tissue...

Page 8: ...rand lens and compensation lens into the light path adjusting center of the Bertrand lens to observe the cone light characteristic of the sample VII MAINTENANCE 1 If there is any dirt settled on the l...

Reviews: