5
STEINDORFF
®
V.
INSTALLATON:
1.
Working environment:
1)
Indoor temperature: 0
℃
~40
℃
, the maximum relative humidity: 85%.
2)
The microscope will be mildewed or damaged in the environment of high temperature
and damp surrounding.
3)
Don’t put microscope in the dusty environment. Please cover it with ethene plastic bag,
when it is not in use.
4)
Please put microscope in the plate place.
2.
Transmission adapter:
Put the transmission adapter on microscope, when screw is connect with orientation slot of the
adapter, tighten the fixing screw with finger.
3.
Binocular or trinocular head:
Put the binocular or trinocular head onto the transmission adapter and tighten with the fixing
screw with finger.
4.
Eyepiece:
Insert eyepiece into the viewing tube.
5.
Objective:
Make the mechanical stage down in order to install the objectives into the nosepiece from the
lowest magnification to the highest in the clockwise direction, please must be install them tightly
in place.
6.
Condenser:
Make sure the oriented screw of the condenser holder is in the proper place.
7.
Complementary color equipment:
Insert complementary color equipment into the proper place in need.
8.
Filter:
It must be put on illuminator according to the user’s needs.
9.
Attachable mechanical stage:
Put attachable mechanical stage on the rotatable mechanical stage.
10.
Adjusting rod:
It is used to adjust center of the objective.