Myo Kinisi Technical Manual
28
Fitting the Hand Shell
19
If the hand shell requires fitting, the instructions below must be followed:
Steeper recommend securing the hand firmly and carefully in a vertical position,
with the hand in a slightly open position and the power supply isolated.
The use of talcum powder is not recommended.
The PVC hand shell can be gently heated to aid the donning process; place the
hand shell on to a clean flat surface in a fan assisted oven at 110°C (230°F) for
90 seconds.
When the hand shell is warm and malleable; carefully ease the wrist section over
the fingers before pulling over the chassis.
Once partially donned, the inner hand shell should be pushed down fully into
place so that there is no bridging and the finger/thumb tips are fully located.
The retaining ring should be mounted to secure the hand shell in place.
If in doubt, please contact Steeper Customer Services.