STATUS SCIENTIFIC CONTROLS
FGD3 Infrared Gas Detector Heads
TD06/024
Issue:
7
Change Note:
1904
Page 20
5.2.8
C: 8 - Restore
The firmware for the detector head is common to both CO
2
and HC instruments. This
feature allows the type of sensor fitted to be selected.
IMPORTANT
The instrument supplied with this manual has either an infrared HC sensor or an
infrared CO
2
sensor installed.
Changing the setting within this menu will not change the gas to which the
sensor is sensitive.
This menu option allows the sensor type to be selected. The available sensor options
are listed below:
OPTION
TYPE
RANGE
COMMENTS
CH4L
Infrared
0-100%LEL
Methane
CH4t
Infrared
0-100%LEL
Methane
CH4H
Infrared
0-100%Vol
Methane
HC
Infrared
0-100%LEL
General hydro carbons
HCtC
Infrared
0-100%LEL
General hydro carbons
HCC
Infrared
0-100% LEL
General hydro carbons
HHC
Infrared
0-100% Vol
General hydro carbons
HHCC
Infrared
0-100% Vol
General hydro carbons
CO2L
Infrared
0-5%Vol
Carbon Dioxide
CO2t
Infrared
0-5%Vol
Carbon Dioxide
CO2H
Infrared
0-25%Vol
Carbon Dioxide
CO2P
Infrared
0-10000 ppm
Carbon Dioxide
ACET
Infrared
0-100%LEL
Acetone
The sensor type is selected as follows:
•
From the menu system select menu option:
C: 8
and press ENTER.
•
Using the NEXT and PREVIOUS buttons, select the required sensor.
•
Press ENTER.
Note:
Pressing ENTER whilst in menu option
C: 8
will reset the factory
defaults for the displayed sensor. A calibration must be performed even
if the sensor type was not changed.
Fault message ‘F 6’ will be
displayed.
•
Press MENU to close the menu system.
Note:
It is important that the sensor selected via this menu option corresponds to the sensor
that is installed. Incorrect settings here may result in the inability of the instrument to
detect gas.