5.2 Menu Mode Selection
The FGD3 uses a password system to restrict the end user from carrying out certain
changes that may compromise the use of the equipment.
If the password system is enabled then the user only has access to calibrate the
sensor and set the 4 - 20 mA without a password.
If the password system is disabled or the Engineer password is entered then complete
access is allowed which gives the user access to every aspect of the system to be
changed.
Note :
The password is a pre-set six digit number which can be obtained from
Status Scientific Controls Ltd.
Several calibration modes exist in the FGD3 and these are accessible via the
instruments simple menu system. To select a calibration mode follow this procedure:
•
Press the MENU button, either
PASS
or
C: 1
appears on the display. If
PASS
appears on the display then either press MENU again or the 6-
digit password.
•
Press UP or DOWN until the required menu option is displayed.
•
Press ENTER to select the calibration mode.
•
To exit the menu press MENU.
While the instrument is in menu mode
– any data displayed on the screen will alternate