Stat Fax 303 Plus Operator'S Manual Download Page 38

34 

Stat Fax 303 Plus Operator’s Manual Rev. J

 

Awareness Technology, Inc. 

2.4 Special Features 

2.4.1 Clock and Calendar 

 To set the correct time and date, use the ALT key to access the command for setting the clock. 
When you press ALT, you will first see 

“SAVE TEST Y/N”

. Press NO. Then you will see 

“STRIP 

TYPE Y/N”

. Again, press the NO key. Now you will see 

“DELETE TEST Y/N”

. Answer NO. 

Then you will see 

“SET CLOCK Y/N”

. This time, press the YES key. 

Depending upon the selected date format, the display will ask for the month, the day, and the 
year (last 2 digits), by displaying 

“Date: MM.DD.YY”

, or the day, month, and year by displaying 

“DD.MM.YY”

. Use two digits for each entry and separate the entries using the decimal point key. 

When the proper date is displayed, press ENTER.  

The instrument will then display 

“Time: HH.MM.SS”

, asking you to enter the hour (1-24), 

minutes, and seconds. Again, use two digits for each entry, and separate your entries using the 
decimal point key. When entering hour, use 13 for 1PM, 14 for 2PM, etc. Press ENTER when the 
displayed time is correct.  

You may configure the reader to display the date either as day.month.year or as month/day/year. 
To select a date format, first press the MENU key and then enter test #100. The display will 
indicate the use of the 0 key to pick the MM/DD format and the use of the 1 key to select the 
DD.MM format. Use the 1 or 0 key to choose and then press ENTER. The display will return to 

“Ready”

Once set, the time and date will not be lost when the instrument is turned off or unplugged. The 
time will be displayed whenever the instrument displays 

“Ready”

. The time and date will be 

printed on all assay reports on either printer. The date last modified will be stored with each test 
in the menu. 

2.4.2 Lamp Saver

To prolong the lamp life, the instrument is programmed to turn off the lamp automatically after a 
few minutes of idle time. The lamp automatically comes on when you enter a mode, or press 
ENTER to continue reading in a mode in which the lamp has timed out. You may also press 

“AUX”

 to turn lamp on or off. Pressing the ENTER key when 

“Set Carrier”

 is in the display will 

cause the lamp to turn on and warm up. After the lamp warms up for 45 seconds, the reader will 
proceed to operate in the previous mode with no need to re-blank or recalibrate.  

2.4.3 Flags and Error Messages 

 Flags are printed messages or labels used to alert the operator of special conditions such as 
approaching limitations. After printing the warnings, the instrument will often continue to perform. 
It is the operator's responsibility to assess any implications. Error messages appear on the 
display when the instrument fails to continue to operate. These messages help the operator 
locate the cause of the problem since it must be overcome before proceeding. The Microstrip 
Reader employs these flags and error messages: 

“>3.00”

 is reported in the absorbance field if a reading exceeds 3A.  

“>10**6”

 is printed as the concentration value whenever the calculated result is greater than 6 

digits and can not be properly printed in the concentration field. 

“CONTROLS FULL”

 is displayed if a fourth control location is attempted. 

Summary of Contents for 303 Plus

Page 1: ...Copyright 2001 Doc 303 Revised 04 2007 Rev J Stat Fax 303 Plus Operator s Manual...

Page 2: ......

Page 3: ...g a Strip Format 16 2 1 2 Selecting a Mode 17 2 1 3 Quitting a Mode 18 2 1 4 Selecting Filters 18 2 1 5 Selecting a Blank 19 2 1 6 Selecting Duplicates 19 2 1 7 Unit Codes 20 2 1 8 Selecting Cutoff In...

Page 4: ...bration and Linearity 36 3 0 Additional Tips and Information 37 3 1 Restoring Electronic Calibration 37 3 2 Operating Precautions 37 3 3 Maintenance 38 3 4 Troubleshooting 39 3 5 References 41 4 0 App...

Page 5: ...ates concentrations and prints results It comes with a built in thermal graphics printer with the capability to plot standard curves Besides providing absorbance readings this instrument also offers p...

Page 6: ...e design with a timed lamp saving feature further assures the reliability of this instrument 1 2 Principles of Operation Light energy from an overhead lamp is focused by a lens directed through an ape...

Page 7: ......

Page 8: ...strument may be impaired Use Proper Power Cord Use only the power cord specified for this product and certified for the country of use Ground the Product This product is grounded through the grounding...

Page 9: ......

Page 10: ...the unit should be turned off Look at the rear panel of the instrument to check that the power switch is in the Off position Power Cord Requirements Use only the power cord specified for this product...

Page 11: ...normally if the power supply is interrupted If this occurs turn the instrument off for a moment When you turn the instrument back on it will resume normal operation but a standard curve which was not...

Page 12: ...cut a new straight edge before trying again Drop the roll of paper into the printer paper well in back spooling up any loose paper Replace the paper cover by pinching its sides together and placing it...

Page 13: ...crostrip reader More details on the operation of each feature are provided in section 2 Operating Procedures The voltage select switch is on the base The power switch and serial connector are on the r...

Page 14: ...ptake Mode Pressing the INTRP key causes the reader to ask for positive and negative cutoff points to be used for automatic sample interpretation Pressing the CNTRL key causes the reader to ask for th...

Page 15: ...total number of wells to read and print The key assignments for the filters are as follows For all models key 1 is 405nm key 2 is 450nm key 3 is 492nm key 4 is 630nm 6 filter UV key 5 is 545nm key 6 i...

Page 16: ...CD 16 characters 5x7 dot matrix Internal Printer Thermal dot matrix 20 characters per line Keyboard 16 key membrane switch 4X4 Calculation Modes Single point calibration Multi point calibration with r...

Page 17: ...supplier Humidity 80 for temperature up to 31 C decreasing linearity to 50 humidity at 40 C Mains supply voltage fluctuations not to exceed 10 of nominal voltage Recommended Environmental Conditions...

Page 18: ...14 Stat Fax 303 Plus Operator s Manual Rev J Awareness Technology Inc Notes...

Page 19: ...o aid the user this instrument may be operated very simply in most modes by bypassing optional features For every test the instrument will require a mode selection and filter choices Thereafter only t...

Page 20: ...lections If the lamp warm up is not completed by the time you press the ENTER key to read the first strip the display will indicate LAMP WARM UP 00 where 00 represents the number of seconds remaining...

Page 21: ...gle point standard curve passing through the point 0 0 A blank is required in the first well or first two wells if duplicate blank is selected A factor equal to the concentration of the calibrator the...

Page 22: ...entration 2 1 3 Quitting a Mode If a test is to be saved in the user test menu this must be done before canceling the mode See section 2 3 1 for instructions on Storing a User Test To quit a mode pres...

Page 23: ...ank and calibrators When duplicates are chosen the mean absorbance reading is used as the reference absorbance for the blank and calibrator You may also choose to read specimens singly or in duplicate...

Page 24: ...utoff When a single upper cutoff is used any concentration which is equal to or greater than the cutoff will be labeled as positive The letters P and N are shown in the right most column of the printo...

Page 25: ...the 8 well mode the first set of wells right most carrier position will be Strip 1 row A through H If you are in the 12 well mode the first set of wells right most carrier position will be Row A stri...

Page 26: ...eak apart wells in the 12 well carrier start from the front of row A and fill all 12 locations then begin at the front of row B etc and continue until done Ensure that the wells are firmly placed in t...

Page 27: ...y for the next strip the display will indicate Set Carrier Then Press Enter To continue reading in the Absorbance mode position the carrier for the next strip and press the ENTER key Since there are n...

Page 28: ...can have any positive or negative numerical value including 0 and 1 Press YES After the filters are selected if you chose no differential the Offset Absorbance questions will follow see section 2 4 5...

Page 29: ...w mNeg C mNeg C Type in each value and press ENTER The display will show Num of PCs Type in the number of Positive Controls and press ENTER The display will show Pos C Pos C Type in each value and pre...

Page 30: ...same order described in the regular cutoff mode NOTE Be sure to read the and signs in each display because blank positive control and negative control parameters are entered in a reverse way when usin...

Page 31: ...e in the number of Cutoff Controls and press ENTER The display will show C O C C O C Type in each value and press ENTER The display will show Num of PCs Type in the number of Positive Controls and pre...

Page 32: ...to access the regression mode Regression Y N will be displayed Answer YES This mode calculates a best fit linear equation based upon the standard points It is designed for use with either linear loga...

Page 33: ...right most carrier position and press ENTER While the strip is being automatically positioned for reading the display will indicate Positioning Strp After the strip or strips containing the standards...

Page 34: ...est 2 calibrator points The calibrators absorbances must be either decreasing or increasing down the strip i e if the first calibrator is the darkest the second must be the next darkest etc and if the...

Page 35: ...e is obtained or the mode is canceled To read and calculate specimens answer NO to the question DELETE WELLS Y N The instrument will be ready for the next strip when it displays Set Carrier Then Press...

Page 36: ...Enter To continue reading in the Uptake mode load the next strip and press the ENTER key Press CLEAR twice to cancel the mode 2 3 User Test Menu This instrument offers the operator 8K bytes of non vol...

Page 37: ...u will see DELETE TEST Y N Answer YES When you see DELETE TEST in the display again type in the number of the test you wish to delete and press the ENTER key The instrument will display DELETE TEST Y...

Page 38: ...e lost when the instrument is turned off or unplugged The time will be displayed whenever the instrument displays Ready The time and date will be printed on all assay reports on either printer The dat...

Page 39: ...ORY IS FULL is displayed if you attempt to store more tests than are available DO ABS SET TEST 212 is printed when the factory set absorbance gain calibration has been lost from non volatile memory In...

Page 40: ...e from your instrument supplier This is because in vertical photometry the fill volume and degree of meniscus determine the pathlength and absorbance is proportional to pathlength For lab test results...

Page 41: ...ires only minimal adjustment from the keyboard to remain calibrated Note that the number entered in test 212 is multiplied by the current absorbance factor set to 1 when reset Thus running test 212 an...

Page 42: ...ive foam This reader is essentially a maintenance free instrument To ensure maximum trouble free operation the instrument need only be kept dry Although the filters are well sealed their life may be d...

Page 43: ...on the internal printer After each well is read the result will remain on the display to permit the user to note the result Press ENTER to advance the strip to the next well Problem The lamp does not...

Page 44: ...oblem The printing is dim or incomplete Solution Use only the specified paper available from your dealer Problem The normal operation of the instrument is suddenly interrupted Solution As with all mic...

Page 45: ...rference filter temperature and humidity recommendations adapted from Melles Griot Optical Guide 3 1985 pp 266 269 2 Other engineering data supplied by Awareness Technology Inc Palm City FL 1987 2006...

Page 46: ...42 Stat Fax 303 Plus Operator s Manual Rev J Awareness Technology Inc Notes...

Page 47: ...rument Appendix 2 is a sheet which may be used to document your stored menu tests Appendix 3 is a set of sample printouts of data from each calculation mode Some modes have been set up to show various...

Page 48: ...d water wait mix and read Using predispensed dyes reduces the degree of pipetting precision required to reconstitute the check set strips Since the strip reader looks through the sample from top to bo...

Page 49: ...Awareness Technology Inc Stat Fax 303 Plus Operator s Manual Rev J 45 4 2 Appendix 2 Test Log Microstrip Reader SN ________ User Test Menu Log Page ___ Test Analyte Stored By Date Notes...

Page 50: ...46 Stat Fax 303 Plus Operator s Manual Rev J Awareness Technology Inc 4 3 Appendix 3 Sample Printouts...

Page 51: ...Awareness Technology Inc Stat Fax 303 Plus Operator s Manual Rev J 47...

Page 52: ...he problem support staff at Awareness Technology is happy to assist you and can be reached in the United States by Phone 772 283 6540 Fax 772 283 8020 e mail support awaretech com Mailing address Awar...

Reviews: