6. Installation Instructions
The Starlab Dry Bath Heating Systems are ready to ‘Plug and Play’. The unit should only be
used on a sturdy, level surface in a safe, dry place. Before operation, ensure one or two heating
aluminium block(s) is / are inserted into the bath, or if using as a water bath, that water has
been added.
7. Operating Instructions
1.
Place the unit on a sturdy, level surface in a safe, dry place, away from laboratory traffic.
2.
Ensure that the AC power switch is OFF, then plug the power cord into a grounded AC outlet of
appropriate voltage (115 V or 220 V as indicated on the rating sticker near the AC cord on the back of
the unit).
3.
Select a suitable metal block(s), or the appropriate water volume if using as a water bath,
and put into the unit.
4.
Turn the AC power ON.
5.
Use the UP or DOWN Key to adjust to the desired temperature.
6.
Press the START Key to start heating.
7.
To change the temperature during heating, press START again to deactivate and set the
new temperature. Press START to activate heating.
8.
Press the START key again, to stop the unit.
NOTE: If setting the
heating time
is required, press the TIMER Key, and then press the UP or
DOWN Key to adjust the time accordingly. The unit will ‘auto-stop’ when the time is up.
Protection/Error Messages
Error Types
Description
Err1
If the displayed temperature value is 5 ºC over the set temperature value,
Err1 will be displayed, an alarm will sound and the instrument will shut down
automatically.
Err2
If the displayed temperature value reaches a temperature much higher than
the set temperature value within 1 minute of the unit being switched on,
Err2 will be displayed and an alarm will sound. This problem is due to SSR.
Err3
If the internal temperature sensor is broken or a connection problem occurs,
Err3 will be displayed; an alarm will sound, and the instrument will shut down
automatically.