7
3. Product Description
The single block (N2400-4001) and dual block (N2400-4002) units use interchangeable heating block
modules for a variety of applications, including restriction digests, denaturing DNA, BUN, melting agar,
coagulation studies, in-situ hybridization and Hot Start thermo-cycler reaction, as well as having the ad-
ditional function of being used as a water bath. All models incorporate a PID controller for easy tempera-
ture selection, rapid heat-up, and excellent stability. Temperature may be set in 0.1 ºC increments, from
5 ºC above ambient to 150 ºC. The timer can be set from 0 to 999 minutes for the user’s convenience.
Features
• Digital performance at analog prices
• Microprocessor controller with digital display
• Easy, user calibration
• Leak-proof aluminium heating chamber
• Single and dual block modes
2. Specifications
Controller
Digital Microprocessor Controller
Display
Digital
Operating Temperature
5 ºC above ambient, to 150 ºC
Temperature Increase
0.1 ºC
Temperature Calibration
Yes
Temperature Uniformity
± 0.2 ºC at 37 ºC
Temperature Accuracy
± 0.2 ºC at 37 ºC
Timer
0 to 999 minutes
Safety Device
Leak-proof heating chamber
Heating Chamber Material
PTFE-coated aluminium
Block Material
Aluminium
Block Dimensions
79 x 104 x 50mm (except N2400-4108)
Water Bath Volume (approx.)
Single: 400 ml. Dual: 800 ml
Data Log
RS 232 (Maximum 2.5 m long)
Power
N2400-4001: 125 W; N2400-4002: 200 W
Rated Voltages
110 or 220 Volts. N2400-4001: 220 V AC, 50/60 Hz, 0.63 A.
N2400-4002: 220 V AC, 50/60 Hz, 1 A.
Unit Dimensions
200 x 298 x 80 mm (w x l x h)
Weight
N2400-4001: 2.1 kg; N2400-4002: 2.8 kg