Voice options
(Figure 13)
• To silence the local alarm without voice message, push switches 3 &
4 to the right. The nurse call cord
must be
connected
(Option A)
.
• For local alarm without voice message, push switch 3 to the left and
switch 4 to the right
(Option B)
.
• For local alarm with voice message, move both switches 3 and 4 to
the left
(Option C)
.
• If you want the voice message to repeat itself, move the number 3
rocker switch to the right and the number 4 rocker switch to the left
(Option D)
. The nurse call cord must be connected for the message
to repeat.
Testing the TABS
®
monitor
1. Before each use, test your TABS
®
mobility monitor.
2. Test the system by pressing firmly on the pressure pad. The monitor
will “chirp” when weight is first applied to the pad to indicate the
monitor is armed. After applying constant pressure for at least 3
seconds, remove the pressure and the alarm will sound. Reset the
alarm by pressing the
RESET
button on the TABS
®
monitor.. For
more instructions on how to use the pads, refer to sections
“
Instructions for chair pressure pad use”
and
“
Instructions for
bed pressure pad use”
for your pad type.
15
Test your TABS
®
mobility monitor and pressure pad before each use to
ensure that it is working properly. Failure to do so may result in injury
or death to a person in your care.
Figure 13
Option A
Option B
Option C
Option D
WARNING
Local and voice message off
(nurse call cord connected)
Voice message off
Voice and local
alarm on
Repeated voice message
(nurse call connected)