3 – 20
Remote Operation
10. The token mode to
OFF
.
*RST
does
not
a
ff
ect
PSTA
,
CONS
,
TERM
, and all service-enable
registers (
*SRE
,
*ESE
,
CESE
,
RCSE
, or
OLSE
).
*RST
Example:
CONS?
1
Identify
*IDN?
Query the device identification string.
The identification string is formatted as:
Stanford Research Systems,SIM918,s/n******,ver#.###
where
SIM918
is the model number,
******
is a 6-digit serial number,
and
#.###
is the firmware revision level.
*IDN?
Example:
Stanford Research Systems,SIM918,s/n005432,ver2.1
Self Test
*TST?
There is no internal self-test in the SIM918 after the power-on, so this
query always returns
0
.
*TST?
Example:
0
Operation Complete
*OPC(?)
Sets the OPC flag in the ESR register.
The query form
*OPC?
writes a
1
into the output queue when com-
plete, but does not a
ff
ect the ESR register.
*OPC?
Example:
1
Console Mode
CONS(?)
{
z
}
Set (query) the console mode
{
to
z
=
(
OFF 0
,
ON 1
)
}
.
CONS
causes each character received at the input bu
ff
er to be copied
to the output queue.
A reset does not alter
CONS
. The value in boldface above is the
power-on value.
CONS
is set to
OFF
upon Device Clear.
CONS ON
Example:
SIM918
Precision Current Preamplifier
Summary of Contents for SIM918
Page 4: ...ii Contents SIM918 Precision Current Preamplifier ...
Page 7: ...General Information v SIM918 Precision Current Preamplifier ...
Page 12: ...x General Information SIM918 Precision Current Preamplifier ...
Page 62: ...3 30 Remote Operation SIM918 Precision Current Preamplifier ...
Page 63: ...4 Circuit Description In This Chapter 4 1 Schematic Diagrams 4 2 4 1 ...