Device embolization
Instructions for Use
Personnel Training
Implanting physicians are required to have successfully completed additional training in
the use of the PA Sensor and Delivery System prior to implant.
Accessories
The accessories required to implant the device and set the sensor’s PA pressure baseline
are listed in the following table. These accessories are not packaged with the device.
Table 4. Accessories
Item
Specifications
Vascular Access Kit
12 Fr Introducer sheath and dilators with access guidewire
PA Catheter
110 cm length
Sensor Delivery
Guidewire
0.018” x 260-300 cm fixed core guidewire with straight or
angled tip (no J-tip)
In addition to the specified accessories, the following catheter lab equipment and supplies
are required to implant and set the sensor’s PA pressure baseline:
•
Fluoroscope with digital angiography capabilities and ability to record and recall
images (C-arm or fixed)
•
Blood pressure monitoring equipment for a right heart catheterization procedure
•
Saline solution
•
Radiopaque contrast media
Package Inspection
Inspect the package carefully before opening and check the Use By date on the product
label. Implant of the sensor is not recommended after its expiration date. If the integrity of
the sterile package has been compromised, or the product or package is defective, do not
use the product and contact Technical Support.
Package Contents
The sensor is packaged separately and supplied sterile. Packages contain:
•
1 PA Sensor and Delivery System
•
1 USB flash drive
•
1 temporary patient implant card
•
Product documentation
Sterilization
The package contents have been sterilized with ethylene oxide before shipment. The
system is for single use and is not intended to be resterilized. If the sterile package has
been compromised, contact Technical Support.
9