Setting Menu

F. Firmware Version : Display the meter's firmware vversion.
G. Temperture

: Sets the unit of measurement for temperture to either Farrenheit(°F)

or Celsius (°C)

H. Stabilization Mode : Sets the meter's clecrification period setting to either Auto and Fixed

Stabilization.

Auto - Enables a 15-second electrification period when the mersured

resistance is 1 x 10

10

Ohms or greater to maitain test accuracy.

Fixed - Complies with ANSI/ESD S4.1 and enables a 15-second

electrification period when the mearured resistance is 1 x 10

6

Ohms or greater.

I. Beep

: Enables and disables the sudible beep when the meter's pushbuttons

are pressed.

J. Erase all memory

: Erase all stored measurement transactions saved in the meter's

memory.

K. Memory Slot Number : Indicates the memory slot number.
L. Resistance Measurement : Indicates the resistance measurement value for the respsctive memory

slot

F

J

K

G

H

I

L

Summary of Contents for MEG102

Page 1: ... can be referred to in the future Measuring Standard Resisatnce Measurement ESD TR53 Work surface ANSI ESD S4 1 Floors ANSI ESD 7 1 Foot Grounder ESD SP9 2 Garments ANSI ESD STM2 1 Seating ANSI ESD STM12 1 Floor Footware with person ANSI ESD STM97 1 Features 1 With Wide range and it is suitable for many material check 2 Easy operation with AUTO range control 3 With several optional probe it makes ...

Page 2: ...mell or sound or smoke or high temperature happens immediately turn off the power and call to our office or local distributor Do not drop or any impact to the product Do not remove any label of product out Without label it will be sometime out of warranty Do not do anything with the device that is not described in this manual Maintenance Clean the tip of the probe with IPA before you use the parti...

Page 3: ... to the power the meter ON Black pushbuttons Each black pushbutton corresponds to the prompts on the bottom left and bottm right of the display These buttons are used to access the Settings and Memory Recall menus and scroll p and down between menu options Red pushbutton Corresponds to the prompts located in the bottom center of the display This button is used to perform tests and select menu opti...

Page 4: ... spacer 254mm 10 914mm 36 4 OMEG RING Optional Probe OMEG B Optional OMEG RING probe is special probe for ANSi ESD S541 ANSI ESD11 11 standard OPTIONAL PROBE With flat conductive metal plate you can measure the resistivity of flat materials For varous shapes you can measure the resistivity with Two small pins Before you start to measure be sure about the contact pins to be straight ...

Page 5: ...g area 3 Each probes must be hold with the materials unless to be incorrect measuring 4 While measuring sometime it happens unstable result with Electrical Noize For such case please re start the power again for normal operation 5 It is danger to cose the product or prove to the objec with big charge it happens directly discharge and sometime damaged to this tester Before measuring ...

Page 6: ...Voltage Displays the test voltage used to complete the measurement C Resistance Measurement Displays the resistance measurement in ohms Ω D Tempareture Dispays the ambient tempareture E Rlative Humidity Displays the relative humidity 1 Opent the battery cover and set the AA type battery and put the cover NOTE Check the battery level Setup the Battery and Probe A B C D E ...

Page 7: ...esistance is 1 x 1010 Ohms or greater to maitain test accuracy Fixed Complies with ANSI ESD S4 1 and enables a 15 second electrification period when the mearured resistance is 1 x 106 Ohms or greater I Beep Enables and disables the sudible beep when the meter s pushbuttons are pressed J Erase all memory Erase all stored measurement transactions saved in the meter s memory K Memory Slot Number Indi...

Page 8: ...t commonly used area of a surface Most worn area Center of surface Furtherst area from a ground point If the suface to be measured has sections i e floor tiles garment panels place the 2 27kg 5lbs electrodes on different sections for Rtt measurements Clean the materials s surface for thest lab measurements but do not clean the surface for materials that are already installed Only clean and re test...

Page 9: ...trode on the most commonly used or worn area The tempareture inside the testing area needs to be around 24 1 7 Temperature at 40 60 relative Humidity The test equipements need to stay under such enviroment around 1 hour for proper readings The meter cannot be noemalized inside objects enclosed boxed containers or cased The temperature inside an eclosed case wil differ from the ourside temperture T...

Page 10: ...nd affect the meter s accuracy at high resistance The frequency of cleaning this area at least onece a month Cable jackets should also be cleaned in this fashion In case the device does not operate correctly it may be the result of one of the following The display pnel is no any indication when the power is turned on Check the battery installation first No battery correct polarity battery life Mea...

Page 11: ...m OMEG B Optional to Test Jacks Φ3 2 37 26 74 5 6 4 Contact Pins with spring MEG102 main body 3 3 4 5 6 7 8 9 10 11 12 12 DIGITAL SURFACE RESISTANCE METER MEG102 32 100 210 Test Jacks Red Push Bottun OLED Display Exponents LEDs Black Push Bottons OMEG RING Optional ...

Page 12: ...HISHIDO ELECTROSTATIC Malfunctions or damage occurring to the product due to fire natural disasters or other acts of providence Other malfunctions or damage occurring to the product deemed not to be the responsibility of SHISHIDO ELECTROSTATIC For any queries relating to the product contact the sales office where you purchased the product SHISHIDO ELECTROSTATIC LTD http www shishido esd co jp engl...

Reviews: