14
1 Getting Started
1.2.3 Lock to External Frequency
In this example the
Ext Sync Input
will be used to synchronize the optical
chopper to an external reference.
To perform the steps in this section, in addition to the setup from the
previous section you will need:
• an external frequency source, such as a function generator, or the
reference output of a lock‑in amplifier.
Beginning with the setup described in Section 1.2.1:
1. Press
to turn on the SR542.
2. Connect the external reference signal to the
Ext Sync Input
connec‑
tor on the rear panel using a standard BNC cable (not included).
3. Press
Source
→
Int
repeatedly to until
Ext Sync
is selected as the source.
4. Press
Edge
repeatedly to select the desired edge triggering. Select
or
for a TTL reference, or
for a sinusoidal reference.
For TTL inputs, a minimum >2V logic level is required. For sinu‑
soidal inputs, an amplitude of at least 100 mV
rms
and frequency
greater than
∼
1 Hz is required.
The
Locked
indicator should illuminate within about three peri‑
ods of the reference frequency. If
Locked
does not light or does
not remain lit, check the quality of the external signal.
5. Use
and
to show the
Source
frequency monitor on
the numeric display and verify that the frequency matches that of
the external reference.
6. Configure the
Control
track by pressing
Control
Shaft
until the desired
track is selected. To select
Shaft
, press and hold
Control
Shaft
.
7. Press and hold
Run /
Stop
until the motor starts.
The
Locked
indicator should remain lit and the
Chopper Locked
indicator will begin to flash indicating the motor has achieved the
correct speed, and then light solid once the phase is locked.
8. Change the frequency of the external source by a few percent.
The
Locked
light will go out indicating that the internal clock is
no longer locked to the external frequency. Until the internal clock
can lock to the new frequency, the chopper will continue to run at
the previous frequency and
Chopper Locked
may remain lit.
Once
both
Locked
and
Chopper Locked
indicators are lit, the chop‑
per is properly phase locked to the external reference.
SR542 Precision Optical Chopper