4-12 Windows Software Error Codes
SRS QMS Gas Analyzer
Troubleshooting:
Check the voltage output of the external power supply with a voltmeter.
Adjust the voltage to 24 V in adjustable external supplies or replace the power
supply altogether if necessary.
Contact SRS for units with a built-in power supply (Option Opt02).
PS7
Type of Error:
24VDC P/S.
Error Message:
External 24V P/S error: Voltage >26V.
Error Cause:
Voltage output of 24VDC Power Supply exceeds the acceptable 22-26V DC
range. The Power LED is turned off and the Error LED is turned on instead.
Troubleshooting:
Check the voltage output of the external power supply with a voltmeter.
Adjust the voltage to 24 V in adjustable external supplies or replace the power
supply altogether if necessary.
Contact SRS for units with a built-in power supply (Option Opt02).
RF4
Type of Error:
Quadrupole Mass Filter RF P/S.
Error Message:
RF P/S Error: Power Supply in current limited mode.
Error Cause:
The circuit that drives the primary of the RF Transformer is in current limited
mode.
Troubleshooting:
Check for a short in the quadrupole connections.
RF6
Type of Error:
Quadrupole Mass Filter RF P/S.
Error Message:
RF P/S Error: Primary Current exceeds 2.0 Amps.