System Problems
4-3
Technical Reference
Loud Noises While Shutting Down
It is common to hear loud noises while the turbo pump is coasting to a stop. The pump is a vibration
source which is slowly scans the frequency range from 1250 Hz to 0 Hz. Any resonances in the chassis
are momentarily excited as the pump passes through the resonant frequency. These noises are normal
and of no concern. Venting the system will cause the turbo pump to quickly decelerate and avoid the
noise.
Nothing Operates
If the main power switch does not light, either the main fuse is blown or the light in the switch is burned
out. The light in the switch is intended to last forever, so this is not the leading suspect.
It the power switch light is on and the fan is working, then the 24 VDC power supply is operational. If
the light is on and the fan is not operating, then the fuse in the 24 VDC power supply could be blown.