Remote/link connector
Neutrik NC5-MAH/FAH
Connectors are used for a link operation of LV unit or for an additional digital remote CMC-xx-DIGI
connection. Up to 30 units can be linked. They are then controlled via one GO and E-STOP button in
the local operation mode.
Pin
Function
note
1
Data CMN
Data Common
2
Data -
Data Minus
3
Data+
Data Plus
4
DC1
Power supply for
CMC
DC12-36V
5
DC2
Power supply for
CMC
DC12-36V
Loadcell E-STOP connector
Neutrik NC4-FAH
Pin
Function
note
1
DC24-36V
Connected to 3
2
Active 1
Active line 1
3
DC24-36V
Connected to 1
4
Active2
Active line 2
Both safety lines are separate and NO /normally open/.
Load-cell E-STOP activation contacts must be in NC /normally closed/ state for E-STOP activation.
To reset the load-cell E-STOP function, please cycle the E-STOP mushroom on the device.