TROUBLESHOOTING
SYMPTOM
REASON/CORRECTIVE ACTION
The characters only appear on half of the
display.
Press the
“WEIGH”
button or disconnect power.
Wait five seconds, then re-connect the power and try
the
“WEIGH”
button again.
The display lights appear to work, but do
not respond to button activation.
Check to ensure the membrane switch label is correctly
plugged into display board.
Check to ensure the calibration switch is not in the
Calibration Mode (Figure 4).
The display shows no reading at all.
Check that wall power supply is plugged in. Check
display cable to make sure it is connected securely.