SYSTEM DESCRIPTION and INTENDED USE 

 
SYSTEM DESCRIPTION: 

The  SRV415  Multi-Purpose  Floor  Scale  has  a  low  profile  design,  which  allows  for  easy  step 
access and can be used in a fixed location or as a portable scale. 

The  SRV415  Multi-Purpose  Floor  Scale  employs  the  latest  in  microprocessor  and  load  cell 
technology  to  provide  accurate  and  repeatable  weight  data.  Four  (4)  identically  matched 
transducers are strategically placed to ensure an accurate representation of weight. 

Weight  is  displayed  on  a  LCD  screen.  With  a  push  of  a  button,  weight  data  may  be  viewed  in 
either

 

pounds or kilograms. 

INTENDED USE: 

The SRV415 Multi-Purpose Floor Scale is designed for table-top use or for stand-on floor use in 
weighing up to a maximum of 500 pounds (227 kilograms). 

 
 
 
 
 
 
 
 
 
 
 

STORAGE and TRANSPORTATION 

 

To maintain proper operation of this instrument, storage and transport conditions should not vary 
outside  the  following  conditions:  Relative  Humidity  0%  to  85%,  Ambient  Temperature  14

F  to 

122

F (-10

C to +50

C). 

 

WARNING 

DO NOT EXCEED 

MAXIMUM WEIGHT LIMIT OF 

500 LB / 227 KG 

DO NOT USE FOR ANIMAL TRANSPORT 

Summary of Contents for SR Scales SRV415

Page 1: ...4 Page 1 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S Instrument...

Page 2: ...nc TABLE OF CONTENTS PACKING CHECKLIST MODEL SRV415 3 ASSEMBLY 3 SYSTEM DESCRIPTION AND INTENDED USE 4 STORAGE AND TRANSPORTATION 4 CLEANING AND DISINFECTING 5 SPECIFICATIONS 6 BUTTON FUNCTIONS 7 BASI...

Page 3: ...otline at 1 800 654 6360 STEP 2 Unlatch tab on the battery compartment cover on bottom of scale Figure 1 Insert batteries according to instructions on the battery compartment holder Replace battery co...

Page 4: ...ogy to provide accurate and repeatable weight data Four 4 identically matched transducers are strategically placed to ensure an accurate representation of weight Weight is displayed on a LCD screen Wi...

Page 5: ...a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available disinfectants...

Page 6: ...olution for calibrated range AUTO ZERO One button operation AUTO POWER DOWN Approximately 60 seconds 1 minute AVERAGING Automatic digital filter POWER SUPPLY Six 6 AA 1 5 volt alkaline batteries CALIB...

Page 7: ...to memory To recall last weight reading press Hold Recall button once scale has shut down LB KG Weight data may be viewed in either pounds or kilograms Pressing the lb kg button allows the operator to...

Page 8: ...reading in memory once the weight is stable If scale cannot lock in reading due to movement press the Hold Recall button to automatically lock the display and store the reading To return to WEIGH mode...

Page 9: ...desired then it will need to be re set Both procedures are found next IMPORTANT CALIBRATION CHECK Qualified service personnel only should perform this procedure Load cells have no user serviceable co...

Page 10: ...2 Full Applied calibrated mass at full scale Half Used only with three point calibration curve Applies calibrated mass at a second point between zero and full scale usually at half zero No calibrated...

Page 11: ...display will show only the five digits XXXX X with the thousand place flashing 1 Select past to the next digit by pressing the Left Button c Once in the hundreds place set the full scale value maximum...

Page 12: ...ply the predetermined full scale value on the scale and press the Left Button e Display will show SAVE press the Left Button to save display will show SAV d or Right Button to toggle to quit Select qu...

Page 13: ...es on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event s...

Page 14: ...o MAN415_200714 Page 14 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S...

Reviews: