SR Instruments SR Scales SRV415 Operating And Service Manual Download Page 11

Model SRV415 Multi-Purpose Floor Scale 
Service Manual 
Part No. MAN415

_

200714 

Page 11 of 14 

 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel: 716-693-5977   Fax: 716-693-5854   URL: 

www.srscales.com

 

email: 

[email protected]

   

 

Copyright 2020

 

S

Instruments, Inc. 

CALIBRATION: (Cont’d) 

 

Gravitational Acceleration (Optional) 

NOTE

:  Setting  the  local  gravitational  acceleration,  “

gLOC

”,  will  not  affect  the 

calibration.  The  local  gravitational  acceleration,  “

gLOC

”,  will  become  the 

calibrated gravitational acceleration, “

gCAL

 

only when it is recalibrated. 

a)

 

Select the Gravitational Acceleration from the CAL menu by pressing the Left Button 
when the display shows “

grav

”. 

b)

 

The display will show only the four digits to the right of the decimal point <9.XXXX>. 

c)

 

Set  the  local  gravitational  acceleration  in  m/s

2

  by  selecting  the  first  digit  and  toggle 

through the digits  using  the Right  Button  until  desired value is  shown. Press  the  Left 
Button and continue the same with the remaining digits. 

d)

 

Press  the  Left  Button  to  save  or  Right  Button  to  toggle  to  "

quit

".  Select  "

quit

"  by 

pressing the Left Button to quit without saving changes. 

e)

 

Press the Left Button again to save and the display will show “

SAV’d

”. 

 

Full-Scale value 

a)

 

Select the Full-Scale from the CAL menu by pressing the Left Button when the display 
shows “

FULL

”. 

b)

 

The  display  will  show  only  the  five  digits  <XXXX.X>  with  the  thousand-place 
flashing <1>. Select past to the next digit by pressing the Left Button. 

c)

 

Once  in  the  hundreds-place,  set  the  full  scale  value  (maximum  weight  capacity)  by 
toggling using the Right Button through the digits until desired value is shown. Select 
the desired value by pressing the Left Button and continue the same with the remaining 
digits. 

d)

 

Press  the  Left  Button  to  save  or  Right  Button  to  toggle  to  "

quit

".  Select  "

quit

"  by 

pressing the Left Button to quit without saving changes. 

e)

 

Press the Left Button again to save and the display will show “

SAV’d

”. 

 

Half-Scale value

 

NOTE

:   The  Half-Scale  value  is  used  only  with  the  three-point  calibration.  It  is  a  value 

between zero and the Full Scale values. It is usually close to half of the Full Scale 
value. 

a)

 

Select the Half-Scale from the CAL menu by pressing the Left Button when the display 
shows “

HALF

”. 

b)

 

The  display  will  show  only  the  five  digits  <XXXX.X>  with  the  thousand-place 
flashing <1>. Select past to the next digit by pressing the Left Button. 

c)

 

Once in the hundreds-place, set the half scale value by toggling using the Right Button 
through the digits until desired value is shown. Select the desired value by pressing the 
Left Button and continue the same with the remaining digits. 

d)

 

Press  the  Left  Button  to  save  or  Right  Button  to  toggle  to  "

quit

".  Select  "

quit

"  by 

pressing the Left Button to quit without saving changes. 

e)

 

Press the Left Button again to save and the display will show “

SAV’d

”. 

Summary of Contents for SR Scales SRV415

Page 1: ...4 Page 1 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S Instrument...

Page 2: ...nc TABLE OF CONTENTS PACKING CHECKLIST MODEL SRV415 3 ASSEMBLY 3 SYSTEM DESCRIPTION AND INTENDED USE 4 STORAGE AND TRANSPORTATION 4 CLEANING AND DISINFECTING 5 SPECIFICATIONS 6 BUTTON FUNCTIONS 7 BASI...

Page 3: ...otline at 1 800 654 6360 STEP 2 Unlatch tab on the battery compartment cover on bottom of scale Figure 1 Insert batteries according to instructions on the battery compartment holder Replace battery co...

Page 4: ...ogy to provide accurate and repeatable weight data Four 4 identically matched transducers are strategically placed to ensure an accurate representation of weight Weight is displayed on a LCD screen Wi...

Page 5: ...a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available disinfectants...

Page 6: ...olution for calibrated range AUTO ZERO One button operation AUTO POWER DOWN Approximately 60 seconds 1 minute AVERAGING Automatic digital filter POWER SUPPLY Six 6 AA 1 5 volt alkaline batteries CALIB...

Page 7: ...to memory To recall last weight reading press Hold Recall button once scale has shut down LB KG Weight data may be viewed in either pounds or kilograms Pressing the lb kg button allows the operator to...

Page 8: ...reading in memory once the weight is stable If scale cannot lock in reading due to movement press the Hold Recall button to automatically lock the display and store the reading To return to WEIGH mode...

Page 9: ...desired then it will need to be re set Both procedures are found next IMPORTANT CALIBRATION CHECK Qualified service personnel only should perform this procedure Load cells have no user serviceable co...

Page 10: ...2 Full Applied calibrated mass at full scale Half Used only with three point calibration curve Applies calibrated mass at a second point between zero and full scale usually at half zero No calibrated...

Page 11: ...display will show only the five digits XXXX X with the thousand place flashing 1 Select past to the next digit by pressing the Left Button c Once in the hundreds place set the full scale value maximum...

Page 12: ...ply the predetermined full scale value on the scale and press the Left Button e Display will show SAVE press the Left Button to save display will show SAV d or Right Button to toggle to quit Select qu...

Page 13: ...es on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event s...

Page 14: ...o MAN415_200714 Page 14 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S...

Reviews: