CALIBRATION: (Cont’d)
Gravitational Acceleration (Optional)
NOTE
: Setting the local gravitational acceleration, “
gLOC
”, will not affect the
calibration. The local gravitational acceleration, “
gLOC
”, will become the
calibrated gravitational acceleration, “
gCAL
”
only when it is recalibrated.
a)
Select the Gravitational Acceleration from the CAL menu by pressing the Left Button
when the display shows “
grav
”.
b)
The display will show only the four digits to the right of the decimal point <9.XXXX>.
c)
Set the local gravitational acceleration in m/s
2
by selecting the first digit and toggle
through the digits using the Right Button until desired value is shown. Press the Left
Button and continue the same with the remaining digits.
d)
Press the Left Button to save or Right Button to toggle to "
quit
". Select "
quit
" by
pressing the Left Button to quit without saving changes.
e)
Press the Left Button again to save and the display will show “
SAV’d
”.
Full-Scale value
a)
Select the Full-Scale from the CAL menu by pressing the Left Button when the display
shows “
FULL
”.
b)
The display will show only the five digits <XXXX.X> with the thousand-place
flashing <1>. Select past to the next digit by pressing the Left Button.
c)
Once in the hundreds-place, set the full scale value (maximum weight capacity) by
toggling using the Right Button through the digits until desired value is shown. Select
the desired value by pressing the Left Button and continue the same with the remaining
digits.
d)
Press the Left Button to save or Right Button to toggle to "
quit
". Select "
quit
" by
pressing the Left Button to quit without saving changes.
e)
Press the Left Button again to save and the display will show “
SAV’d
”.
Half-Scale value
NOTE
: The Half-Scale value is used only with the three-point calibration. It is a value
between zero and the Full Scale values. It is usually close to half of the Full Scale
value.
a)
Select the Half-Scale from the CAL menu by pressing the Left Button when the display
shows “
HALF
”.
b)
The display will show only the five digits <XXXX.X> with the thousand-place
flashing <1>. Select past to the next digit by pressing the Left Button.
c)
Once in the hundreds-place, set the half scale value by toggling using the Right Button
through the digits until desired value is shown. Select the desired value by pressing the
Left Button and continue the same with the remaining digits.
d)
Press the Left Button to save or Right Button to toggle to "
quit
". Select "
quit
" by
pressing the Left Button to quit without saving changes.
e)
Press the Left Button again to save and the display will show “
SAV’d
”.