CALIBRATION: (Cont’d) 

 

ACCESS SYSTEM SETUP

 

To access all settings below, remove the feet, spacers and the seven screws holding on the bottom cover. 
Press the calibration button (Figure 3) and CAL mode is now active. 

The  Right  Button  allows  you  to  toggle  through  the  menus  without  saving  any  parameters.  The  Left 
Button permits a menu item to be selected or a parameter saved (Figure 4)

NOTE

:  Ensure  that  nothing  is  in  contact  with  the  scale  system  during  this  procedure.  Remove 

hands from the system when noting the displayed calibration results. 

CAL MENU 

“data”  

Displays saved calibration and model information. 

“unit”  

Set the available units; lb only, kg only, or selection of lb or kg with mode button. 

“sec”  

Set Auto Power Down Timer between 15 and 900 seconds. 

“grav”  

Set Gravitational Acceleration in m/s

2

“Full” 

Applied calibrated mass at full scale. 

“Half” 

Used only with three point calibration curve. Applies calibrated mass at a second 
point between zero and full scale, usually at half. 

“zero” 

No calibrated weight applied. 

“2 pt” 

Two- point calibration curve, zero and full scale. 

“3 pt” 

Three-point calibration curve; zero, half and full scale. 

“save” 

Saves the calibration data. 

“quit” 

Quits without saving the calibration data. 

 

Enter CAL Mode 

To  enter  the  CAL  mode,  press

 the 

calibration  button

 (

Figure  3

). Toggle through the menu 

with Right Button (

Figure 4

). Select or Save data with the Left Button (

Figure 4

).

 

 

Auto Power Down Timer (Optional)

 

a)

 

Select the Auto Power Down Timer from the CAL menu by pressing the Left Button 
when the display shows “

sec

”. 

b)

 

The display will show four digits <0123>. 

c)

 

Set the desired time in seconds by selecting the first digit and toggle through the digits 
using the Right Button until desired value is shown. Press the Left Button and continue 
the  same  with  the  remaining  digits.  The  value  cannot  be  less  than  15  or  greater  than 
900. 

d)

 

Press  the  Left  Button  to  save  or  Right  Button  to  toggle  to  "

quit

".  Select  "

quit

"  by 

pressing the Left Button to quit without saving changes. 

e)

 

Press the Left Button again to save and the display will show “

SAV’d

”. 

 

Summary of Contents for SR Scales SRV415

Page 1: ...4 Page 1 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S Instrument...

Page 2: ...nc TABLE OF CONTENTS PACKING CHECKLIST MODEL SRV415 3 ASSEMBLY 3 SYSTEM DESCRIPTION AND INTENDED USE 4 STORAGE AND TRANSPORTATION 4 CLEANING AND DISINFECTING 5 SPECIFICATIONS 6 BUTTON FUNCTIONS 7 BASI...

Page 3: ...otline at 1 800 654 6360 STEP 2 Unlatch tab on the battery compartment cover on bottom of scale Figure 1 Insert batteries according to instructions on the battery compartment holder Replace battery co...

Page 4: ...ogy to provide accurate and repeatable weight data Four 4 identically matched transducers are strategically placed to ensure an accurate representation of weight Weight is displayed on a LCD screen Wi...

Page 5: ...a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available disinfectants...

Page 6: ...olution for calibrated range AUTO ZERO One button operation AUTO POWER DOWN Approximately 60 seconds 1 minute AVERAGING Automatic digital filter POWER SUPPLY Six 6 AA 1 5 volt alkaline batteries CALIB...

Page 7: ...to memory To recall last weight reading press Hold Recall button once scale has shut down LB KG Weight data may be viewed in either pounds or kilograms Pressing the lb kg button allows the operator to...

Page 8: ...reading in memory once the weight is stable If scale cannot lock in reading due to movement press the Hold Recall button to automatically lock the display and store the reading To return to WEIGH mode...

Page 9: ...desired then it will need to be re set Both procedures are found next IMPORTANT CALIBRATION CHECK Qualified service personnel only should perform this procedure Load cells have no user serviceable co...

Page 10: ...2 Full Applied calibrated mass at full scale Half Used only with three point calibration curve Applies calibrated mass at a second point between zero and full scale usually at half zero No calibrated...

Page 11: ...display will show only the five digits XXXX X with the thousand place flashing 1 Select past to the next digit by pressing the Left Button c Once in the hundreds place set the full scale value maximum...

Page 12: ...ply the predetermined full scale value on the scale and press the Left Button e Display will show SAVE press the Left Button to save display will show SAV d or Right Button to toggle to quit Select qu...

Page 13: ...es on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event s...

Page 14: ...o MAN415_200714 Page 14 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S...

Reviews: