Model SR725
i
Wheelchair Scale
Operating and Service Manual
Part No. MAN725
i
_MICROBAN_200603
Page 13 of 24
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
email:
Copyright 2020
S
Instruments, Inc.
BUTTON FUNCTIONS
Figure 8: Button Display
ZERO / WEIGH
Press and hold to zero. The display will read “
WT = 0.0 Lb
”. This is used to zero the
system before placing the patient on the scale system. This action also resets previously
stored weight, height and BMI values to zero. Ensure that nothing is in contact with the
weighing surface during this procedure.
Press to weigh. Weight stable indicator “
” flashes then remains solid when stable. Auto
stores stable weight in memory.
SEND (PRINTER / EHR)
Press to send stored values to EHR or to printer. Output values include time, date, weight,
height, and BMI. Also, indicated on display when paper is out, “
PAPER OUT
” and when
door is open, “
PAPER DOOR OPEN
”.
RECALL
Press to recall the last stable weight. Display will scroll thru “
WT
”, “
HT
”, and “
BMI
”
stored values in approximately two second intervals.
MENU
Press Menu to edit setup.
Setting
the
UNITS
: Use
UP
or
DOWN
arrow buttons to select “
Lb
” or “
Kg
”. Press
ENTER
to save changes.
Setting
ON TIME
: Use
UP
or
DOWN
arrow buttons to adjust the “
ON TIME
”. The
“
ON TIME
” may be set from 30 to 300 seconds in 30 second increments. Press
ENTER
to save changes.
Setting
TIME
and
DATE
: Use the
UP
arrow button to select digit. To change digit use
the
DOWN
arrow button. Press
ENTER
to save changes.
NOTE
: When selected, the year position defaults to “00”
Continued next page