Model SR585
i
Bariatric Stand-On Scale
Operating and Service Manual
Part No. MAN585
i
-KG_MICROBAN_201006
Page 12 of 22
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
www.srscales.com
email:
Copyright 2020
S
Instruments, Inc.
BUTTON FUNCTIONS (Cont’d)
MENU
Press Menu to edit setup.
Setting
ON TIME
: Use
UP
or
DOWN
arrow buttons to adjust the “
ON TIME
”. The
“
ON TIME
” may be set from 30 to 300 seconds in 30 second increments. Press
ENTER
to save changes.
Setting
TIME
and
DATE
: Use the
UP
arrow button to select digit. To change digit use
the
DOWN
arrow button. Press
ENTER
to save changes.
NOTE
: When selected, the year position defaults to “00”
Configuring scale output: Use the
UP
or
DOWN
arrow buttons to select between
and
EHR.
Press
ENTER
to save changes.
Setting printout orientation: Use the
UP
or
DOWN
arrow buttons to select between
“
PRINT INVERTED
” and “
PRINT NORMAL
”. Press
ENTER
to save changes.
UP
Press
UP
to adjust height up from the default, to increase the scale’s “on time”, or to
select a digit when setting time and date.
DOWN
Press
DOWN
to adjust the height down from the default, to decrease the scale’s “on
time”, or to change the value of a selected digit when setting time and date.
BASIC SYSTEM OPERATION
SETTING SYSTEM ZERO
Make sure the scale is free and clear of any obstructions and press and hold the
ZERO /
WEIGH
button. The displayed message will indicate
“
HOLD TO ZERO
” and count
down to zero.
Make sure that nothing is in contact with the scale while zeroing the system.
In a few seconds, the display will read “
WT = 0.0
Kg
. This action also resets previously
stored weight, height, and BMI values to zero.
When using AC Power Supply
: The scale requires only an initial setting of System Zero,
or any time the display reads “
PLEASE ZERO
”. The scale will otherwise zero track and
remove any slight fluctuations.
When using batteries for power:
It is recommended that the system be zeroed prior to
each use.
Continued next page