OPERATING MANUAL
MN-SST70
ED-19-01
SPRIANO® TECHNOLOGIES
- www.spriano.it – [email protected]
a TERRANOVA® Srl brand
- www.terranova-instruments.com - VAT IT07848810151
Factory & Sales: Via Gramsci 1 – 26827 Terranova dei Passerini (LO) – Italy
Head Office: Via Rosso Medardo 16 – 20159 Milano – Italy Ph: +39 0377 919119 – Fax: +39 0377 855720
Pa
g.
20
/52
Figure 15: earthing in the housing
W
ARNING
!
The hose must not be considered as a protection against electric shock for the
protection of security, but as protection against electromagnetic interference of
the instrument. Therefore, in the case of long distance cable (L> 20 m), both ends
of the braid must be connected to the chassis or to the connectors. In the case of
cables L <20 meters, will be sufficient to connect the braid only from the side
“Power Supply”.
Figure 16: earthing of the shield for cables longer than 20 meters