Electromagnetic Compatibility
This product conforms to the EMC standard (IEC 60601-1-2:2007).
•
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided in the accompanying documents.
•
Connectors should not be touched and connections should not be made to these connectors unless ESD
precautionary procedures are used.
•
ESD protection mark may be used in lieu of protective covers.
•
Portable and mobile RF communications equipment can effect medical electrical equipment.
•
The use of accessories and cables other than those supplied with the instrument exception for cables sold
by the manufacturer of the equipment as replacement parts, may result in increase emission or decrease
immunity of the equipment or system.
•
The equipment or system should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the equipment or system should be observed to verify normal operation in the
configuration in which it will be used.
EMC Table
Emission Guidance
The Pediavision Spot is intended for use in the electromagnetic environment specified below. The operator or the user of Spot should assure that
it is used in such an environment.
Emission Test
Compliance
Electromagnetic environment guide
RF Emission Cispr 11
Group 1
Spot must emit electromagnetic energy in
order to perform its intended function.
Nearby electronic equipment may be affected.
RF Emission Cispr 11
Class A
Spot is suitable for use in all establishments
other than domestic and those directly
connected to the public low-voltage power
supplies building used for domestic purposes.
Harmonic Emission IEC 61000-3-2
Class A
Compliant
Voltage Fluctuation/flicker Emission IEC
61000-3-3