www.spirit-healthcare.co.uk
57
Influence Quantities
Packed cell volume (Hematocrit)
Packed cell volume evaluation was conducted in various
hematocrit levels. The range of hematocrit levels within the
acceptance criteria is 15–65 %.
Interferences
The effect of various interfering substances was evaluated in
whole blood samples. The presence of the following substances
within the given concentrations does not affect blood glucose
measurements. Higher concentrations of the substances shown
below may cause inaccurate blood glucose results.
No.
Interferent
Concentration
1
Acetaminophen (paracetamol)
20 mg/dL
2
Ascorbic acid
3 mg/dL
3
Bilirubin
20 mg/dL
4
Ceftriazone
81 mg/dL
5
Cholesterol
500 mg/dL
6
Creatinine
3 mg/dL
7
Dopamine
13 mg/dL
8
EDTA
180 mg/dL
9
Galactose
60 mg/dL
10
Gentisic acid
50 mg/dL
11
Glutathione (Red)
17 mg/dL
12
Hemoglobin
500 mg/dL
13
Heparin
8000 U/dL