56
www.i-sens.com
Within ±0.83 mmol/L and Within ±15 %
599/600 (99.8 %)
System accuracy results for glucose concentrations between
2.2 mmol/L and 26.0 mmol/L
Precision:
The precision studies were performed in a laboratory
using TEE2 Plus Blood Glucose Monitoring Systems.
Within Run Precision
Blood average
2.3 mmol/L
SD = 0.1 mmol/L
4.7 mmol/L
SD = 0.2 mmol/L
6.9 mmol/L
CV = 3.4 %
11.4 mmol/L
CV = 3.3 %
17.8 mmol/L
CV = 3.5 %
Between Run Precision
Control average
2.2 mmol/L
SD = 0.2 mmol/L
6.7 mmol/L
CV = 2.5 %
18.8 mmol/L
CV = 4.0 %
This study shows that there could be variation of up to 4.0 %.