Operating Manual
6
1300-87075-1A
Installation
The unit should be installed on a sturdy bench to avoid any vibration during operation. When locating the unit,
there should be enough space allowed for the drawer to move out without obstruction.
Intended Use
The Genoseal is intended to be used for the heat sealing of microplates or tubes during experimentation,
storage, or transportation. This device is not intended to be used as a medical device.
Caution:
Use appropriate lifting and carrying equipment available in the facility during unit movement
from one place to another. Always use any necessary PPE during unit movement.
Inventory
Note:
All of the items listed below are available as spare components from Spex.
DESCRIPTION
PART NUMBER
Plate Adapter for PCR 384-Well Format
2060-PA-384
Skirtless Plate Adapter
2060-PA-96
Heat Sealing Film
2061
Plate Weight for Recessed Plates*
2060-WT-RC
Plate Weight for Standard and Rimless Plates*
2060-WT-STD
Plate Locators*
2060-PL
* Accessories supplied with the Genoseal
Operating Environment
The Genoseal should be operated in an environment with a temperature range of 10 °C to 30 °C and a non-
condensing relative humidity range of 10% to 80%.
The unit should be kept out of drafts and air currents, as they will have an adverse effect upon the temperature
stability of the sealing plate area. If the sealing plate area is subjected to drafts or air currents, quality or process
time may be affected.
The unit should be placed on a secure and stable surface with a minimum of 100 mm (4-inch) clearance on all
sides.
Connecting the Power Supply
The Genoseal utilizes a standard IEC inlet, which is fused and switched. Plug the IEC Lead in the inlet and switch
the Rocker I.O. Switch to the On position ( | ). The display screen should illuminate.
Note:
The system is self-regulated to sense voltage differences between 100 / 250 VAC.