background image

adjustment screw clockwise or counterclock-
wise until the needle indicator is directly over
the three black zeros at the left end of the scale.

Analog Meter: Reading the Scale

General

The first step in reading the analog scale is to
align the needle with the scale.

Resistance (Ohms - 

Ω

)

Use the top scale for reading the resistance.
If the meter is set to X1k, multiply the
resistance value by 1000

Ω

.

DC Voltage (V DC)

Use the middle scale. Match the dial setting to
the highest number on the scale.

AC Voltage (V AC)

Use the same numbers and procedures as
used for the DC voltage setting, but use the
scale directly below the numbers.

DC Milliamps

Use the same scale and procedure as used
for the DC voltage setting.

DC

1

2

5

10

20

40

100

200

1K

6

8

}

30

60

90

120

150

60

120

180

240

300

0

5

4

8

7

Decibel Gain (dB)

- see pg. 15

Use the scale marked dB to read decibels for
transistor gain. Use the chart at the right of
the scale for proper conversion.

3.2 Internal Battery Condition

Before making resistance or continuity tests,
check the condition of the internal battery. First
turn the function/range switch to the ohms
Rx1K position. Short the test leads together
and the needle indicator should deflect to the
right side of the scale. Keep the test leads
shorted together while simultaneously turning
the zero ohms adjustment dial until the needle
indicator reads zero at the right side of the
ohms (green) scale. If the needle will not zero,
replace the battery with a new 1.5 volt AA size
non-rechargeable battery (see 

Battery

Replacement).

4. DC Voltage Measurement

1) Fully seat the test leads in the correct input

jacks, (-) black lead, (+) red lead.

2) Set the function/range switch to the 

appropriate DC voltage range. If the voltage
is unknown, use the highest range. If the 
voltage applied falls within the range of a
lower setting, reset the function/range switch
to the appropriate setting for greater accuracy.

3) If the polarity of the circuit to be tested is

known, touch the black test lead to the 
neutral side. If the polarity is unknown, touch
the test leads to opposite sides of the circuit.
If the needle indicator deflects to the left of
the scale, reverse the test leads.

Resistance
(Ohms)
Scale

DC Voltage 
& Milliamps
Scale

Common Markings 
for both the AC & DC
Voltage Scales

AC Voltage Scale

dB Scale for Transistor 

Gain Measurement

HSP5 3-09 manual.qxp  5/8/09  9:57 AM  Page 8

Summary of Contents for GMT-312

Page 1: ...ual Read this owners manual thoroughly before use and save HSP5 HSP5 3 09 manual qxp 5 8 09 9 57 AM Page 1 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at...

Page 2: ...or the homeowner hobbyist and professional needing to make electrical and electronic equipment measurements 1 1 Meter Functions Needle Analog Scale Mechanical Zero Adjust Screw Zero Ohms Adjust ment I...

Page 3: ...m max Decibels 20dB to 56dB on AC voltage ranges Accuracy DC voltage mA 4 full scale of range AC voltage 5 full scale of range Resistance 4 arc of scale length Sensitivity 2000 ohms per volt DC and AC...

Page 4: ...et the function range switch to WARNING WARNING 6 5 the appropriate setting for greater accuracy 2 Avoid placing the meter in areas where vibration dust or dirt are present Do not store the meter in e...

Page 5: ...n range switch to the ohms Rx1K position Short the test leads together and the needle indicator should deflect to the right side of the scale Keep the test leads shorted together while simultaneously...

Page 6: ...1 Common DC Milliamperage Measurements It is important to point out that milliamps can also be expressed as thousandths of an Ampere therefore 150 milliamps is 150 thousandths of one Amp The 150mA fun...

Page 7: ...nce and continuity testing POWER MUST BE OFF 1 Fully seat the test leads in the input jacks black lead red lead 2 Set the function range switch to the Rx1K position resistance indicated multiplied by...

Page 8: ...fuse is good the needle indicator will move to 0 ohms If not replace the fuse On plug type fuses touch the the test leads on the bottom contact and the other on the threaded metal contact see fig 6 7...

Page 9: ...uses GB model GF 0306 4 Carefully replace the back cover and tighten the screw Do not overtighten as this may strip the threads in the meter housing 10 Warranty 1 YEAR WARRANTY limited solely to repai...