Sper scientific 850007 Instruction Manual Download Page 7

7

Sampling Time 

The sampling time is the time allotted between successive 

measurements. To adjust the sampling time (in seconds):
1.  Enter the sampling time function from Setup Mode (see page 5). 

 

“SP-t” appears on the LCD. 

2.  Press ▲ or ▼ to adjust the value (0, 1, 2, 5, 10, 30, 60, 120, 300, 

 

600, 800, 1800, 3600 seconds). Press 

ENTER

 to save the value. 

SD Memory Card Format

Enabling this function will format the SD memory card to work 

specifically with your meter. Formatting the SD card will erase any 

previous memory on the card. New SD cards should be formatted 

 

to work with your meter. 
1.  Enter the SD memory card format function from Setup Mode 

 

(see page 5). “Sd F” appears on the LCD. 

2.  Press ▲ or ▼ to select yes (format the SD memory card) or no 

 

(do not format the SD memory card). Press 

ENTER

 to confirm your 

selection. If selecting yes, “yES Enter” will appear on the LCD, 

 

press 

ENTER

 again and the meter will format the SD card. 

MEASUREMENT PROCEDURES

Turning the Unit On/Off

Press 

POWER

 to turn the meter on. 

Press and hold 

POWER

 for 2 seconds to turn the meter off. 

Selecting the Function

1.  Press and hold 

FUNC 

to cycle through the options listed below. 

Release 

FUNC

 when you reach the desired function. 

Light Meter: “LIgHt” appears on the LCD.
Type K/J thermometer: “tP” appears on the LCD.

Note…

 

The meter will default to the last function setting used when it is  

turned off and on again. 

Summary of Contents for 850007

Page 1: ...Visible Light SD Card Datalogger 850007 Instruction Maual...

Page 2: ...hanical or other means that do not yet exist or may be developed including photocopying recording or any information storage and retrieval system without the express permission from Sper Scientific IN...

Page 3: ...nt time and date FEATURES Functions as a light meter and Type K J thermometer Three ranges for light measurement Auto range for light measurement Lux or Foot Candles unit selection SD memory card data...

Page 4: ...pter the meter will remain permanently ON and the POWER button will be disabled METER COMPONENTS KEYPAD 1 2 3 4 5 6 12 11 7 8 9 10 13 1 Type K J Thermocouple Socket 2 Light Sensor Input Socket 3 LCD S...

Page 5: ...k the meter will automatically enter the decimal type setup function 1 Press POWER to turn the meter on 2 Press SET for 2 seconds or longer to enter Setup Mode 3 Press NEXT to cycle through the setup...

Page 6: ...m Setup Mode see page 5 PoFF appears on the LCD 2 Press or to select yes auto power off enabled or no auto power off disabled Press ENTER to save the selection Touch Tone 1 Enter the touch tone functi...

Page 7: ...he SD memory card format function from Setup Mode see page 5 Sd F appears on the LCD 2 Press or to select yes format the SD memory card or no do not format the SD memory card Press ENTER to confirm yo...

Page 8: ...ght sensor zero adjustment is needed 1 With the sensor cover on press ZERO for longer than 3 seconds A value of 0 will appear on the LCD 2 Remove the sensor cover to resume normal measurement Type K J...

Page 9: ...The meter will return to Normal Mode Note The meter cannot be turned off from the memory record function Exit the function then press and hold POWER to turn the meter off Backlight The backlight turns...

Page 10: ...alogging pause the datalogger Press REC for 2 secs or longer REC will disappear from the LCD to indicate that datalogging has ended Note To enable disable the touch tone feature see page 6 Manual Data...

Page 11: ...the datalogger and executing the function again the data is saved to the LXA01001 XLS file until the data reach 30 000 data columns A new file will then be generated i e LXA01002 XLS 3 The folder LXA...

Page 12: ...meter 3 Remove the old batteries and replace with six new AA batteries ensuring correct polarity 4 Replace the battery cover Tighten the screws on the battery cover to secure to the meter TROUBLESHOOT...

Page 13: ...Sampling Time of Display Approximately 1 second Data Output RS232 USB PC computer interface Memory Card SD card 1 GB to 16 GB Operating Temperature 0 to 50 C Operating Humidity 85 RH Power Supply Alk...

Page 14: ...Lux Foot Candles Auto Range 200 Ft cd 0 to 186 Ft cd 2 000 Ft cd 167 to 1 860 Ft cd 10 000 Ft cd 1 670 to 9 290 7 Ft cd Range Resolution Accuracy 2 000 Lux 1 Lux 4 reading 2 dgt 20 000 Lux 10 Lux 100...

Page 15: ...1300 C 50 1 to 100 C 0 1 C 0 4 reading 0 5 C 0 4 reading 1 C 58 0 to 2372 F 58 1 to 148 F 0 1 F 0 4 reading 1 F 0 4 reading 1 8 F Type J 50 0 to 1200 C 50 1 to 100 C 0 1 C 0 4 reading 0 5 C 0 4 readi...

Page 16: ...y does not cover probes batteries battery leakage or damage resulting from accident tampering misuse or abuse of the product Opening the meter to expose its electronics will break the waterproof seal...

Reviews: