Oxystart® - EN
Guidance and manufacturer’s declaration-electromagnetic immunity
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer’s declaration-electromagnetic immunity
The Oxystart
®
Pulse Oximeter is intended for use in the electromagnetic environment specified below. The
customer or the user of the Oxystart
®
Pulse Oximeter should ensure that it is used in such an environment.
Immunity test
IEC60601 test
level
Compliance
level
Electromagnetic environment - guidance
Radiated RF
ICE 61000-4-3
3V/m
80MHz to
2.5GHz
3 V/m
Portable and mobile RF communication equip-
ment should be used no closer to any part of
the Oxystart
®
Pulse Oximeter,including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
80 MHz to 800 MHz
800 MHz to 2,5 GHz
Where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recom-
mended separation distance in meters (m).
P
E
d
=
1
7
P
E
d
=
1
5
.
3