Power Connections (I/O System Interface Box):
Note: Included cables with the I/O Box not shown for the
Generator Power, or XRVC Controller
Connect the ground lug on the rear panel to earth ground. Use a minimum of #10 AWG copper stranded
wire. Remove the top cover of the I/O Box for the Chiller or Cooler Interlock and power cable
installation. The Chiller or Cooler includes instructions for sizing the input power wire and detailing
water or oil flow and thermal interlock connections. Remove the top of the I/O Box and locate TB-4 and
TB-5. Route the input power wires from the Cooler or Chiller through the “Cooler” wire grommet and
connect to TB-4 Terminals 1, 2 and 3. Make sure that #3 is the ground connection from the Cooler or
Chiller. The Cooler or Chiller voltage is normally 230Vac single phase. Consult the manufacturer for
additional information. Connect the interlock switches for temperature and coolant flow from the
Cooler or Chiller through the “Cooler “grommet to TB-5, Terminals 1 and 3.
Refer to cutout of I/O Box
diagram for connection locations.
Interlock Switches Cooler/Chiller
Summary of Contents for XRV Sub-system
Page 1: ...SPELLMAN HIGH VOLTAGE XRV SUBSYSTEM INSTALLATION MANUAL ...
Page 13: ...Page 13 of 98 118157 001 Rev C XRV 160kV Overview 400Vac 3Ø Input ...
Page 17: ...Page 17 of 98 118157 001 Rev C XRV225kV Overview XRV225 400Vac 3Ø 6kW Unit ...
Page 21: ...Page 21 of 98 118157 001 Rev C XRV450 Overview XRV450 Bipolar Cathode Generator ...
Page 22: ...Page 22 of 98 118157 001 Rev C XRV450 Overview XRV450 Bipolar Anode Generator ...
Page 28: ...Page 28 of 98 118157 001 Rev C XRVC Controller Overview XRVC Controller Rear Panel Detail ...
Page 51: ...Page 51 of 98 118157 001 Rev C System Grounding Detail and HV Cable Connections ...
Page 66: ...Page 66 of 98 118157 001 Rev C ...
Page 67: ...Page 67 of 98 118157 001 Rev C Power Source Connections Unipolar Basic ...
Page 68: ...Page 68 of 98 118157 001 Rev C Power Source Connections Unipolar Basic ...
Page 72: ...Page 72 of 98 118157 001 Rev C Power Connections I O System Interface Box ...
Page 76: ...Page 76 of 98 118157 001 Rev C Main Input Power Connections I O System Interface Box ...
Page 77: ...Page 77 of 98 118157 001 Rev C I O Subsystem Interface Box System Diagram ...
Page 89: ...Page 89 of 98 118157 001 Rev C X Ray Tube Filament Calibration Setup ...