Page
61
of
98
118157-001 Rev C
Power Connections (Unipolar Basic):
Use stranded copper wire #8 AWG for the Input power Mains and Ground connection. Use stranded
copper wire #12 AWG for the Auxiliary Line, Neutral and Ground connection. Input power requirements
for the 4kW Generators are 180Vac to 264Vac ± 10%, 30A maximum for Mains and 5A maximum for
Auxiliary.
The 6 kilowatt generator uses an MS3106A-24-22S style connector for the Mains Input and an
MS3106-20-3S style connector for the Auxiliary Power Input. The 6kW generators require 3 Phase
208Vac ±10%, 30A maximum input or 400Vac ±10%, 15A Max. Input. The input power is selected when
placing the order for the XRV Generator . The Auxiliary Power Input requires a separate 230 Vac ±10%,
5A maximum source. Use #10 AWG copper stranded wire for each phase and ground on the Input Power
Mains JB-1. Use #12 AWG copper stranded wire for each line and ground on the Input Power Auxiliary
JB-2.
6kW Input Power Connectors
Summary of Contents for XRV Sub-system
Page 1: ...SPELLMAN HIGH VOLTAGE XRV SUBSYSTEM INSTALLATION MANUAL ...
Page 13: ...Page 13 of 98 118157 001 Rev C XRV 160kV Overview 400Vac 3Ø Input ...
Page 17: ...Page 17 of 98 118157 001 Rev C XRV225kV Overview XRV225 400Vac 3Ø 6kW Unit ...
Page 21: ...Page 21 of 98 118157 001 Rev C XRV450 Overview XRV450 Bipolar Cathode Generator ...
Page 22: ...Page 22 of 98 118157 001 Rev C XRV450 Overview XRV450 Bipolar Anode Generator ...
Page 28: ...Page 28 of 98 118157 001 Rev C XRVC Controller Overview XRVC Controller Rear Panel Detail ...
Page 51: ...Page 51 of 98 118157 001 Rev C System Grounding Detail and HV Cable Connections ...
Page 66: ...Page 66 of 98 118157 001 Rev C ...
Page 67: ...Page 67 of 98 118157 001 Rev C Power Source Connections Unipolar Basic ...
Page 68: ...Page 68 of 98 118157 001 Rev C Power Source Connections Unipolar Basic ...
Page 72: ...Page 72 of 98 118157 001 Rev C Power Connections I O System Interface Box ...
Page 76: ...Page 76 of 98 118157 001 Rev C Main Input Power Connections I O System Interface Box ...
Page 77: ...Page 77 of 98 118157 001 Rev C I O Subsystem Interface Box System Diagram ...
Page 89: ...Page 89 of 98 118157 001 Rev C X Ray Tube Filament Calibration Setup ...