17
Display Installation:
1.
Follow the procedures for
Display Removal
to access the
display and cable connections.
2.
Re
-
connect the sensor cable to the 3.5mm connector on the
back of the display module.
3.
Insert the foam spacer back into place behind the sensor cable.
The split end fits around the sensor cable molding.
4.
If equipped with the IR temperature sensor option, attach it to
the smaller diameter connector.
5.
Guide the excess cable length back down through the base
plate.
6.
Align the arrows on the base and display in the same
direction. The USB port should face the same side as the
serial number label.
7.
Tighten the four mounting screws.
Sensor Block Removal / Replacement:
1.
Follow steps for
Display Removal
to
access the cable connections.
2.
Remove the thumb screw lock bolt
that joins the lower probe base to the
upper shaft.
3.
Separate the probe base from the shaft
(fig. 2).
4.
Feed the cable from the replacement probe base through the
upper shaft. Attaching a string fed down from the top can aide
in the process of cable reconnection.
5.
Follow the procedure for
Display Installation
to complete the
replacement.
Figure 2. Shaft interface