4
SPECIFICATIONS
Irradiance Display Resolution: 5µW/cm
2
over the entire range
Voltage: All models available in 120V/60Hz, 230V/50Hz or 100V/50-60Hz
+
-
CONTROL PANEL DISPLAY DESCRIPTION
1. Power
— ON/OFF switch. When ‘‘ON,’’ the Spectrolinker electronics are
in the WAKE state before receiving value or operational mode
instructions.
2. Optimal Crosslink
— Sets energy dose to the optimal value of 120
µ
J/cm
2
.
3. Time
— Enter time up to 9,999 seconds. NOTE: The minimum interval for
value entry is 1 second.
4. Energy
— Select energy dose up to 0.9999 J/cm
2
. NOTE: The
minimum interval for value entry is 100
µ
J/cm
2
(0.1 mJ/cm
2
).
5. Intensity
— Measures UV irradiance inside the chamber.
6. Start
— Initiates the selected operation (i.e., ENERGY, TIME, OPTIMAL
CROSSLINK, or INTENSITY). This pad turns on the UV tubes after the
mode and value are entered.
7. Reset
— Clears all previously entered instructions and returns the
Spectrolinker electronics to the WAKE state.
8. Numeral Pads
— Selects desired time or energy level. This must be
done after mode selection.
9. X100
µ
J/cm
2
— Lit LED indicates ‘‘ENERGY’’ or ‘‘OPTIMAL
CROSS LINK’’ mode operation.
10. Seconds
— Lit LED indicates ‘‘TIME’’ mode operation.
11. µW/cm
2
— Lit LED indicates ‘‘INTENSITY’’ mode operation.
12. Display
— Displays selected values of time, energy, or intensity level as
well as the following useful prompts: ‘‘DOOR’’ for open door, ‘‘BULB’’ for
intensity level under 1,500
µ
W/cm
2
, and ‘‘END’’ for completed
operation.
XL-1000
XL-1500
UV Tubes
(5) 8-watt 254nm
(6) 15-watt 254nm
Overall Housing
Dimensions
(W x H x D)
19.5 x 10.5 x 9 in
(49.5 x 26.7 x 22.9 cm)
24 x 10.5 x 14 in
(61.0 x 26.7 x 35.6 cm)
Effective
Inner Chamber
Dimensions
(W x H x D)
13.5 x 7 x 7.5 in
(34.3 x 17.8 x 19.1 cm)
18.25 x 6.25 x 12.5 in
(46.4 x 15.9 x 31.8 cm)
Net Weight
17.5 lb (7.9 kg)
26 lb (11.8 kg)