background image

SpectraPure

SpectraPure®Inc.

  

480.894.5437   Call us toll-free  1.800.685.2783    

215 South Industrial Drive,   Suite 2A   Tempe, Arizona   85281-2941

®

20

Membrane Output Calculation Example

What is the expected GPD from a 75 GPD System at 40 psi pressure and 60°F 
water temperature?
 

PCF = 40 ÷ 60 = 0.666

 

TCF = 0.754 (from Table 1)

 

Expected GPD = 75 × 0.666 × 0.754 = 37.7 GPD ± 15%

Performance Test

The performance of a RO membrane is measured in terms of its rejection charac-
teristics. 

Important: Test the quality of the membrane once every 6 months.

Note: This procedure will require a Conductivity Meter (TS-C61).

Procedure:

 

1.  Measure tap water conductivity. (Call it X)

 

2.  Run the system for 15-20 minutes.

 

3.  Rinse test instrument cell 2-3 times with RO water.

 

4.  Measure RO water conductivity. (Call it Y).

 

5.  Subtract RO water conductivity from tap water conductivity. (X - Y) 

 

6.  Divide this quantity by tap water conductivity. (X - Y) ÷ X 

 

7.  Rejection = [(X - Y) ÷ X ] ×100 

* Conductivity in the above procedure could be replaced by hardness, alkalinity, 
nitrate, phosphate, silica etc. (measured in ppm or mg/l).

Rejection of the RO Membrane Calculation Example

 

1.  Tap water hardness = 150 ppm (X)

 

2.  RO water hardness = 7 ppm (Y)

 

3.  X - Y = 143 ppm

 

4.  (X - Y) ÷ X = 143 ÷ 150 = 0.953

 

5.  Rejection = [ ( X - Y)  ÷ X ] ×100 = 0.953 ×100 = 95.3

 

Membrane Hardness Rejection = 95.3 % :  Membrane OK

Production Rate

SpectraPure

®

SpectraPure®Inc

 Fax  480.894.6109 Fax us toll-free 1.877.527.7873

E-mail: [email protected] Visit us on the web www.spectrapure.com

5

SpectraPure®Inc

. assumes no responsibility for water 

damage due to leaks. It is the user’s responsibility to determine 

that the system is leak-free.

System Start-Up

System Hook-Up

1.  Attach the garden hose adapter 

(black tubing)

 to your cold water 

source.   Never run hot water (greater than 100° F (38° C)) through the 
 system.
2. Place 

the 

yellow concentrate (waste) tubing

 and the 

blue purified 

(product) water tubing

 into a drain. Do not restrict flow from these lines.

3.  Open the cold water supply valve. You may use pressure up 
 

to 80 psi (5.5 bar). 

Check the system to ensure that all fittings are tight and leak-
 

free before leaving the system unattended.

Note: It is recommended that at least 2 gallons (7.57 liters) of puri-
fied water be discarded before collecting purified water for use. If 
the unit is not used for several days, run the system for at least 15 
minutes before collecting any water.

System Initialization

If you are setting up your system for the first time or replacing a 
membrane, please see the “Checking the Concentrate to Purified 
Water Ratio” section on page 6 .

Summary of Contents for The Eliminator DI

Page 1: ...f membrane elements for one year from date of receipt by the buyer providing that the loss of performance was not caused by fouling neglect or water conditions exceeding the feed water parameters list...

Page 2: ...ane 19 20 Membrane Testing System Troubleshooting Guide 21 Membrane Troubleshooting Guide 21 System Care Tips for Long Membrane Life 22 Storage 22 Choosing a Mounting Location 22 System Components Rep...

Page 3: ...Temperature Extremes 1 The unit must be kept out of areas that are subject to freezing temperatures 2 High temperatures greater than 100 F 38 C must be avoided If the unit is used outside avoid puttin...

Page 4: ...Low water temperature Heat feed water or use higher GPD membrane Low feed pressure Use booster pump or use higher GPD membrane Fouled membrane Replace membrane Membrane Troubleshooting Guide The follo...

Page 5: ...3 X Y 143 ppm 4 X Y X 143 150 0 953 5 Rejection X Y X 100 0 953 100 95 3 Membrane Hardness Rejection 95 3 Membrane OK Production Rate SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 87...

Page 6: ...illiliters per Minute 5 If the actual concentrate flow rate is less than the optimum concentrate flow rate you must reduce the length of the flow restrictor capillary tubing as described on the next p...

Page 7: ...ncentrate tubing Fig C for Single Membrane System or Fig D for Dual Membrane System Remove the tubing from its push fitting at the membrane as follows a Firmly depress and hold the push fitting collar...

Page 8: ...ail and remove the blue product water inter connect tubing 9 Lift the dual RO membrane assembly from the holding clips 10 Separate the two membranes by depressing the collar on the push fitting of the...

Page 9: ...mbrane 1 Membrane 2 feed push fitting black inter connect feed tubing blue product water interconnect tubing Membrane 1 concentrate push fitting Membrane 1 product water push fitting SpectraPure Spect...

Page 10: ...raPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 15 Note The flow restrictor assembly should remain uninst...

Page 11: ...as described in the Removal Adjustment and Replacement section on page 7 Snap the membrane assembly back into its holding clips System Maintenance RO Membrane Replacement If you have a Single Membran...

Page 12: ...and check for leaks System Maintenance SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 13 System Mai...

Reviews: