Tips for Long Membrane Life

1.   Replacement of 1 micron sediment filter once every 6 months.This will pre 
 

 vent mem brane fouling due to silt or sediment depositing on the membrane. 

2.   Replacement of 10 micron carbon block filter at least once every 6 months  
 

 or when  chlorine breakthrough occurs . This will ensure good membrane life 

 

 and protect the  membrane from chlorine damage.

3.   Membrane should not be operated at lower than the recommended concen 
 

 trate to  purified water ratios, as described on page 6.

4.   Operating reverse osmosis systems on softened feed water greatly reduces
 

  the  chances of membrane fouling.

5.   Use the optional flush kit valve after each use of the system to extend
 

  membrane life  up to 6 months.

Storage

1.   It is recommended that you store your Eliminator RO System in a cool
 

 place when  not being used.

2.   Your Eliminator RO System must be protected from freezing or temperatures
 

  above  100° F (38°C).

Choosing a Mounting Location

When considering a location for the installation of the System, consider the 
following factors:

  

Light 

Sources

 

1.  Most of the components of this system are plastic and are subject to

 

 damage  by ultra- violet light from the sun and other sources such as metal

 

 halide   lighting.

 

2. Algae is more likely to thrive inside the clear filter housings when exposed to 

  

bright 

light.

 

3.  Avoid installing this unit in bright light or direct sunlight.

  

Temperature 

Extremes

 

1. The unit must be kept out of areas that are subject to freezing temperatures.

 

2. High temperatures greater than 100° F (38° C) must be avoided. If the unit is

 

 used outside, avoid putting the system in direct sunlight or connecting it to a

 

 garden hose that may be exposed to sunlight.

System Care

SpectraPure

®

SpectraPure®Inc

 Fax  480.894.6109 Fax us toll-free 1.877.527.7873

E-mail: [email protected] Visit us on the web www.spectrapure.com

3

System Specifications

Sediment Pre-Filter   

1 micron MicroTec™  sediment pre-filter

Carbon Filter 

 

 

1 micron carbon block pre-filter

RO Membrane Type  

Encapsulated thin-film composite 

DI 

Cartridge 

   

SilicaBuster™ Color-Indicating Mixed-Bed 

Rejection Rate   

 

Greater than 98% average

Input Water Pressure 

60 psi (4.15 bar) line pressure*

Input Water Temp    

77°F (25°C)

Recovery Rate   

 

20% (i.e. 20% of the water will be

       collected 

as 

pure 

water)

Nominal Membrane Flow Rates @ 60 psi,  77° F, & 500 ppm TDS :
 

GPD (lpd)  Product Water Flow Rate 

 

Concentrate Flow Rate   

  

 

90 

(340) 

 235 

ml/min 

    940 

ml/min 

 

180  (680)   

470 ml/min  

 

 

      1880 ml/min

Reverse Osmosis Membrane Feed Water

Requirements

Operating Pressure* 

 

 

40 – 80 psi (2.75 – 5.5 bar)

pH 

Range 

     3 

– 

11

Temperature Range  

 

 

38°F – 100° F (3°C – 38°C)

Maximum 

Turbidity 

   1.0 

NTU

Maximum Silt Density Index   

5.0 (based on 15 min. test time)

Maximum Chlorine   

 

 

less than 0.1 ppm

Maximum 

TDS 

    2000 

ppm

Maximum Hardness  

 

 

10 grains (170 ppm as CaCO3)

Maximum 

Iron 

    less 

than 

0.1 

ppm

Maximum Manganese   

 

less than 0.1 ppm

Maximum Hydrogen Sulfide  

0 ppm

Langlier Saturation Index 

 

LSI must be negative

*Operating pressure less than 40 psi will require a high flow booster 
pump (BPHF-MO-115(230)).
*Operating pressure greater than 80 psi will require a pressure reducing 
valve.

System Information

Summary of Contents for The Eliminator DI

Page 1: ...f membrane elements for one year from date of receipt by the buyer providing that the loss of performance was not caused by fouling neglect or water conditions exceeding the feed water parameters list...

Page 2: ...ane 19 20 Membrane Testing System Troubleshooting Guide 21 Membrane Troubleshooting Guide 21 System Care Tips for Long Membrane Life 22 Storage 22 Choosing a Mounting Location 22 System Components Rep...

Page 3: ...Temperature Extremes 1 The unit must be kept out of areas that are subject to freezing temperatures 2 High temperatures greater than 100 F 38 C must be avoided If the unit is used outside avoid puttin...

Page 4: ...Low water temperature Heat feed water or use higher GPD membrane Low feed pressure Use booster pump or use higher GPD membrane Fouled membrane Replace membrane Membrane Troubleshooting Guide The follo...

Page 5: ...3 X Y 143 ppm 4 X Y X 143 150 0 953 5 Rejection X Y X 100 0 953 100 95 3 Membrane Hardness Rejection 95 3 Membrane OK Production Rate SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 87...

Page 6: ...illiliters per Minute 5 If the actual concentrate flow rate is less than the optimum concentrate flow rate you must reduce the length of the flow restrictor capillary tubing as described on the next p...

Page 7: ...ncentrate tubing Fig C for Single Membrane System or Fig D for Dual Membrane System Remove the tubing from its push fitting at the membrane as follows a Firmly depress and hold the push fitting collar...

Page 8: ...ail and remove the blue product water inter connect tubing 9 Lift the dual RO membrane assembly from the holding clips 10 Separate the two membranes by depressing the collar on the push fitting of the...

Page 9: ...mbrane 1 Membrane 2 feed push fitting black inter connect feed tubing blue product water interconnect tubing Membrane 1 concentrate push fitting Membrane 1 product water push fitting SpectraPure Spect...

Page 10: ...raPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 15 Note The flow restrictor assembly should remain uninst...

Page 11: ...as described in the Removal Adjustment and Replacement section on page 7 Snap the membrane assembly back into its holding clips System Maintenance RO Membrane Replacement If you have a Single Membran...

Page 12: ...and check for leaks System Maintenance SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 13 System Mai...

Reviews: